Mechanical failure of the distal radius volar locking plate.
نویسندگان
چکیده
PURPOSE To review 9 cases of mechanical failure of the volar locking plate for distal radial fractures. METHODS Records of 374 consecutive patients who underwent volar locking plating for distal radial fractures were reviewed. Mechanical failures of the volar locking plate were defined as plate breakage or bending, screw breakage or loosening, or collapse of articular fragments resulting in intra-articular screw extrusion. RESULTS Nine mechanical failures occurred between 2 weeks and 3 months in 8 (2.4%) of the patients aged 25 to 82 (median, 74) years with AO fracture types of A3 (n=4), C1 (n=1), C2 (n=1), and C3 (n=3). Mechanical failures included screw pullout (n=5), locking plate bending (n=2), locking screws breakage (n=1), and loosening of locked variable angle screws (n=1). One patient underwent revision of fixation and 2 underwent implant removal. The remainder were treated conservatively. All patients were followed up for a minimum of 12 months; their mean flexion arc was 87 degrees (standard deviation [SD], 17) and the mean rotation arc was 136 degrees (SD, 29 degrees). According to the Green and O'Brien score, their outcomes were good (n=1), fair (n=4), and poor (n=3). CONCLUSION Although mechanical failure of volar locking plate is uncommon, some are potentially preventable.
منابع مشابه
Volar locking plate fixation versus Kirschner wire fixation in distal radius fractures: a meta-analysis
Objective: The superiority of volar locking plate fixation and K-wire fixation on treatment of distal radius fractures was controversial. Thus, we performed a meta-analysis to compare the efficacy of volar locking plate and K-wires for distal radius fracture. Methods: We searched Embase, Medline and PubMed for randomized controlled trials which compare the effects of volar locking plate and K-w...
متن کاملSuccessful closed reduction of refractured wrist with a bent volar distal radius plate.
A case is presented of a 78-year-old woman who sustained a re-fracture of the distal radius after initial fixation with a volar locking plate. The fracture was located in the mid-portion of the volar plate with palmar apex bending of the plate and signs of median nerve compression. A successful closed reduction was performed without hardware failure. The fracture united uneventfully.
متن کاملFailure of volar locking plate fixation of an extraarticular distal radius fracture: A case report
BACKGROUND Volar locking plates provide significant structural stability to the distal radius. Failure of a volar locked plating is a rarely reported complication in the literature. CASE PRESENTATION A 40 year-old, obese female patient who presented with a displaced extraarticular distal radius fracture, underwent open reduction and internal fixation of the fracture using a volar locking plat...
متن کاملDistal radius nonunion after volar locking plate fixation of a distal radius fracture: a case report
We describe a 59-year-old man who had nonunion of a right distal radius fracture after volar locking plate fixation. He underwent open reduction and internal fixation with a volar locking plate system for a dorsally displaced, unstable distal radius fracture at a previous hospital 5 months ago. Radiographs of the injured wrist showed nonunion of the distal radius with 1.5-mm ulnar minus varianc...
متن کاملBiomechanical comparison of volar locked plate constructs using smooth and threaded locking pegs.
The goal of this study was to determine whether there is any biomechanical difference in terms of construct strength with axial loading between volar fixed-angle locking plates with threaded locking vs smooth locking pegs. The control group comprised 7 cadaveric specimens with threaded locking pegs, and the test group comprised 7 cadaveric specimens from the same donor with smooth locking pegs....
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Journal of orthopaedic surgery
دوره 21 3 شماره
صفحات -
تاریخ انتشار 2013