Percutaneous right atrial pacemaker lead repositioning using a regular deflectable ablation catheter

نویسندگان

  • Erdem Ozel
  • István Osztheimer
  • Ali Ozturk
  • László Gellér
  • Emin Evren Ozcan
چکیده

Lead dislocation is one of the most undesired complications of permanent pacemaker implantation. The rate of dislodgement of atrial pacing leads is nearly 3% [1]. Reopening the pacemaker pocket and repositioning the lead is the most commonly used technique for correcting dislocations which occur in the early period after implantation. However, this increases the risk of pocket infection and increases lead insulation problems. Snare systems and urological baskets have also been used for repositioning atrial leads [2, 3]. We report a patient who experienced dislodgement of the atrial pacemaker lead of a cardiac resynchronization therapy (CRT) device, which was corrected using a regular deflectable ablation catheter by the transfemoral route. A 77-year-old female patient with severe ischemic left ventricular dysfunction and wide QRS had received a CRT device. At the follow-up 1 week after the procedure, atrial sensing and pacing dysfunctions were detected. Fluoroscopy revealed dislodgement of a passive fixation J-shaped atrial lead. Lead repositioning via a percutaneous transfemoral approach was planned to avoid surgical revision. A regular deflectable ablation catheter (steerable 4-mm-tip ablation catheter, Mariner, Medtronic, Minneapolis, USA) was advanced to the right atrium through a 7 Fr introducer from the right femoral vein. We propagated the ablation catheter above the level of the atrial lead and made an anteflexion to the distal portion of the catheter. We captured the distal loop of the atrial lead, then with a slight counterclockwise rotation and pull back we oriented the atrial lead towards the right atrial appendage (Figure 1). The lead functions were normal after the intervention. The patient was discharged the next morning and had normal pacemaker functions at follow-up. Early lead displacements after pacemaker implantations are the most frequent cause of reintervention in the majority of patients. We performed successful atrial lead repositioning with a deflectable ablation catheter by

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Successful percutaneous repositioning of a dislodged atrial pacemaker lead with a deflectable catheter.

A 67-year-old male patient with ischemic dilated cardiomyopathy underwent implantation of an implantable cardioverter-defibrillator with cardiac resynchronization therapy about one month ago. The implantation process was successful and the patient was discharged in good condition. At the routine first month follow-up, pacemaker analysis revealed that there was no atrial pacing and sensing, alth...

متن کامل

Luminal esophageal temperature monitoring with a deflectable esophageal temperature probe and intracardiac echocardiography may reduce esophageal injury during atrial fibrillation ablation procedures: results of a pilot study.

BACKGROUND Luminal esophageal temperature (LET) monitoring is one strategy to minimize esophageal injury during atrial fibrillation ablation procedures. However, esophageal ulceration and fistulas have been reported despite adequate LET monitoring. The objective of this study was to assess a novel approach to LET monitoring with a deflectable LET probe on the rate of esophageal injury in patien...

متن کامل

Catheter ablation of atrial fibrillation without fluoroscopy using intracardiac echocardiography and electroanatomic mapping.

BACKGROUND Catheter ablation of atrial fibrillation is currently guided by x-ray fluoroscopy. The associated radiation risk to patients and medical staff may be significant. We report an atrial fibrillation ablation technique using intracardiac echocardiography (ICE) and electroanatomic mapping without fluoroscopy. METHODS AND RESULTS Twenty-one patients with atrial fibrillation (age, 42 to 7...

متن کامل

Pacemaker lead entrapment complicating transcatheter closure of a patent foramen ovale.

T ranscatheter cardiovascular interventions are nowadays more frequently and safely performed, thus leading to populate the heart with medical devices. Here, we describe an unusual complication following percutaneous closure of a patent foramen ovale (PFO) with an atrial septal occluder. A 52-year-old woman with history of transient ischemic attack was found to have a PFO associated with an atr...

متن کامل

Electrocardiographic Diagnosis of Acute Coronary Syndrome in Patients With Endocavitary Pacemakers

Rev Esp Cardiol. 2009;62(9):1068-81 1079 surgically created.1,2 The norm is, as occurred in this case, the absence of an acP potential in the ablation exit point, due to the fact that these pathways are a consequence of a direct connection between the atrial appendage and the ventricular myocardium. Ablation from the right atrial appendage usually requires its isolation with the applications of...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:

دوره 12  شماره 

صفحات  -

تاریخ انتشار 2016