Adaptive bolus chasing computed tomography angiography: Control scheme and experimental results
نویسندگان
چکیده
In this paper, a new adaptive bolus-chasing control scheme is proposed to synchronize the bolus peak in a patient's vascular system and the imaging aperture of a computed tomography (CT) scanner. The proposed control scheme is theoretically evaluated and experimentally tested on a modified Siemens SOMATOM Volume Zoom CT scanner. The first set of experimental results are reported on bolus-chasing CT angiography using realistic bolus dynamics, real-time CT imaging and adaptive table control with physical vasculature phantoms. The data demonstrate that the proposed control approach tracks the bolus propagation well, and clearly outperforms the constant-speed scheme that is the current clinical standard.
منابع مشابه
Study of an adaptive bolus chasing CT angiography.
To improve imaging quality and to reduce contrast dose and radiation exposure, an adaptive bolus chasing CT angiography was proposed so that the bolus peak position and the imaging aperture can be synchronized. The performance of the proposed adaptive bolus chasing CT angiography was experimentally evaluated based on the actual bolus dynamics. The experimental results show that the controlled t...
متن کاملAdaptive Bolus Chasing Computed Tomography Angiography
Synchronization of the contrast bolus peak and CT imaging aperture is a crucial issue for CTA. It effects the CTA imaging quality and the necessary amount of contrast dose. In this paper, we propose an optimal adaptive bolus chasing controller. The controller estimates and predicts the unknown two dimensional bolus density on line and then determines the optimal control actions. Tracking errors...
متن کاملAdaptive bolus-chasing computed tomography angiography in the cases of symmetric and asymmetric arterial flows in peripheral arteries
Synchronization of the contrast bolus peak and CT imaging aperture is a crucial issue for computed tomography angiography (CTA). It affects the CTA image quality and the amount of contrast dose. A whole-body CTA procedure means to scan from the abdominal aorta to pedal arteries. In this context, the synchronization is much more difficult with the asymmetric arterial flow in lower extremities th...
متن کاملBolus characteristics based on Magnetic Resonance Angiography
BACKGROUND A detailed contrast bolus propagation model is essential for optimizing bolus-chasing Computed Tomography Angiography (CTA). Bolus characteristics were studied using bolus-timing datasets from Magnetic Resonance Angiography (MRA) for adaptive controller design and validation. METHODS MRA bolus-timing datasets of the aorta in thirty patients were analyzed by a program developed with...
متن کاملA Dynamic Arterial Tree Phantom for studies of bolus chasing CT Angiography
Computed Tomography (CT) and Magnetic Resonance (MR) imaging techniques have become an important role in diagnosing the vascular diseases. Before a clinical trial of the newly developed techniques, it is essential to evaluate them on a phantom. However, there is no existing arterial A Dynamic Arterial Tree Phantom for studies 89 tree phantom that serves for flood flows. To that end, we design a...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Biomedical signal processing and control
دوره 3 4 شماره
صفحات -
تاریخ انتشار 2008