Using the ACR CT accreditation phantom for routine image quality assurance on both CT and CBCT imaging systems in a radiotherapy environment
نویسندگان
چکیده
Image-guided radiation therapy using cone-beam computed tomography (CBCT) is becoming routine practice in modern radiation therapy. The purpose of this work was to develop an imaging QA program for CT and CBCT units in our department, based on the American College of Radiology (ACR) CT accreditation phantom. The phantom has four testing modules, permitting one to test CT number accuracy, slice width, low contrast resolution, image uniformity, in-plane distance accuracy, and high-contrast resolution reproducibly with suggested window/levels for image analysis. Additional tests for contrast-to-noise ratio (CNR) and noise were added using the polyethylene and acrylic plugs. Baseline values were obtained from CT simulator images acquired on a Phillips Brilliance Big Bore CT simulator and CBCT images acquired on three Varian CBCTs for the imaging protocols most used clinically. Images were then acquired quarterly over a period of two years. Images were exported via DICOM and analyzed manually using OsiriX. Baseline values were used to ensure that image quality remained consistent quarterly, and baselines were reset at any major maintenance or recalibration. Analysis of CT simulator images showed that image quality was within ACR guidelines for all tested scanning protocols. All three CBCT systems were unable to distinguish the low-contrast resolution plugs and had the same high-contrast resolution over all imaging protocols. Analysis of CBCT results over time determined a range of values that could be used to establish quantitative tolerance levels for image quality deterioration. While appropriate for the helical CT, the ACR phantom and guidelines could be modified to be more useful in evaluating CBCT systems. In addition, the observed values for the CT simulator were well within ACR tolerances.
منابع مشابه
Erratum: Using the ACR CT accreditation phantom for routine image quality assurance on both CT and CBCT imaging systems in a radiotherapy environment
-
متن کاملEvaluation of the gray level in CBCT systems and its relationship with HU in CT Scanners
Introduction: Cone-beam CT (CBCT) is an imaging system which offers three-dimensional (3D), multiplanar images and has many advantages over computed tomography (CT) including shorter acquisition times for the resolution desired in dentistry, lower radiation dose to the patient, reasonable price and higher spatial resolution but CBCT scanners are unable to display actual Hounsf...
متن کاملA feasibility study on the use of MV-CBCT images for urgent palliative treatment planning
Introduction: The application of 3D volumetric imaging modalities in treatment planning of radiation therapy can provide more precisely define tumor localization, and computed tomography (CT) is the most common accepted method for treatment planning. Given the lack of a CT scanner stationed in all radiotherapy departments and equipping most of the medical linear accelerators wi...
متن کاملDosimetric Study of an Indigenous and Heterogeneous Pelvic Phantom for Radiotherapy Quality Assurance
Introduction: In vitro dosimetric verification prior to patient treatment plays a key role in accurate and precision radiotherapy treatment delivery. Since the human body is a heterogeneous medium, the aim of this study was to design a heterogeneous pelvic phantom for radiotherapy quality assurance. Material and Methods: A pelvic phantom was ...
متن کاملMethods to evaluate the performance of kilovoltage cone-beam computed tomography in the three-dimensional reconstruction space
Background: Cone-beam computed tomography (CBCT) scanners for image-guided radiotherapy are in clinical use today, but there has been no consensus on uniform acceptance to verify the CBCT image quality yet. The present work proposed new methods to fully evaluate the performance of CBCT in its three-dimensional (3D) reconstruction space. Materials and Methods: Compared to the traditional methods...
متن کامل