mARC vs. IMRT radiotherapy of the prostate with flat and flattening-filter-free beam energies
نویسندگان
چکیده
BACKGROUND There as yet exists no systematic planning study investigating the novel mARC rotational radiotherapy technique, which is conceptually different from VMAT. We therefore present a planning study for prostate cancer, comparing mARC with IMRT treatment at the same linear accelerator equipped with flat and flattening-filter-free (FFF) photon energies. METHODS We retrospectively re-contoured and re-planned treatment plans for 10 consecutive prostate cancer patients. Plans were created for a Siemens Artiste linear accelerator with flat 6 MV and FFF 7 MV photons, using the Prowess Panther treatment planning system. mARC and IMRT plans were compared with each other considering indices for plan quality and dose to organs at risk. All plans were exported to the machine and irradiated while measuring scattered dose by thermoluminescent dosimeters placed on an anthropomorphic phantom. Treatment times were also measured and compared. RESULTS All plans were found acceptable for treatment. There was no marked preference for either technique or energy from the point of view of target coverage and dose to organs at risk. Scattered dose was significantly decreased by the use of FFF energies. While mARC and IMRT plans were of very similar overall quality, treatment time could be markedly decreased both by the use of mARC and FFF energy. CONCLUSIONS Highly conformal treatment plans could be created both by the use of flat 6 MV and FFF 7 MV energy, using IMRT or mARC. For all practical purposes, the FFF 7 MV energy and mARC plans are acceptable for treatment, a combination of both allowing a drastic reduction in treatment time from over 5 minutes to about half this value.
منابع مشابه
Dosimetric Comparison between 6MV Flattened Filter and Flattening Filter Free Photon Beams in the Treatment of Glioblastoma with IMRT Technique: A Treatment Planning Study
Introduction: The present study evaluated the dosimetric comparison between 6MV flattened filter (FF) and flattening filter-free (FFF) photon beams in intensity-modulated radiation therapy (IMRT) technique for the treatment of glioblastoma (GBM) patients. Material and Methods: The present study was conducted on 10 patients with GBM previously...
متن کاملmARC Treatment of Hypopharynx Carcinoma with Flat and Flattening-Filter-Free Beam Energies – A Planning Study
BACKGROUND The recently implemented mARC-rotation-technique is capable to deliver high dose rate bursts. For the case of hypopharynx cancer plans we evaluate whether the mARC can achieve an advantage in treatment time in comparison to IMRT. These plans consider two arcs with flat and flattening filter free (FFF) beam energies. MATERIALS AND METHODS For 8 hypopharynx-cancer patients step-and-s...
متن کاملDosimetric influence of Flattening Filter (FF) and Flattening Filter Free (FFF) 6 and 10 MV photon beams on Volumetric Modulated Arc Therapy (VMAT) planning in case of prostate carcinoma
Background: In the treatment of prostate cancer, radiotherapy is the potential to increase second primary cancers such as bladder and rectal cancers. The reasons for this potential are more monitor units (MUs), therefore a larger total body dose because of leakage radiation, a bigger volume of normal tissue is exposed to lower radiation doses. This study was designed to compare the integral dos...
متن کاملComparative Planning of Flattening-Filter-Free and Flat Beam IMRT for Hypopharynx Cancer as a Function of Beam and Segment Number
Although highly conformal dose distributions can be achieved by IMRT planning, this often requires a large number of segments or beams, resulting in increased treatment times. While flattening-filter-free beams offer a higher dose rate, even more segments may be required to create homogeneous target coverage. Therefore, it is worthwhile to systematically investigate the dependence of plan quali...
متن کاملEnhancement of the Dose on 12 MV Linac with Free Flattening Filter Mode
Purpose: In the last years, some studies investigated dosimetric benefits of a free flattening filter for the photon mode in the radiotherapy field. This study aims to provide a theoretical study published and analysis of basic dosimetric properties for a Saturne 43 Linac head to implement free flattening filter beams clinically. Material and Methods: Thi...
متن کامل