Preclinical multimodality phantom design for quality assurance of tumor size measurement
نویسندگان
چکیده
BACKGROUND Evaluation of changes in tumor size from images acquired by ultrasound (US), computed tomography (CT) or magnetic resonance imaging (MRI) is a common measure of cancer chemotherapy efficacy. Tumor size measurement based on either the World Health Organization (WHO) criteria or the Response Evaluation Criteria in Solid Tumors (RECIST) is the only imaging biomarker for anti-cancer drug testing presently approved by the United States Food and Drug Administration (FDA). The aim of this paper was to design and test a quality assurance phantom with the capability of monitoring tumor size changes with multiple preclinical imaging scanners (US, CT and MRI) in order to facilitate preclinical anti-cancer drug testing. METHODS Three phantoms (Gammex/UTHSCSA Mark 1, Gammex/UTHSCSA Mark 2 and UTHSCSA multimodality tumor measurement phantom) containing tumor-simulating test objects were designed and constructed. All three phantoms were scanned in US, CT and MRI devices. The size of test objects in the phantoms was measured from the US, CT and MRI images. RECIST, WHO and volume analyses were performed. RESULTS The smaller phantom size, simplified design and better test object CT contrast of the UTHSCSA multimodality tumor measurement phantom allowed scanning of the phantom in preclinical US, CT and MRI scanners compared with only limited preclinical scanning capability of Mark 1 and Mark 2 phantoms. For all imaging modalities, RECIST and WHO errors were reduced for UTHSCSA multimodality tumor measurement phantom (≤1.69 ± 0.33%) compared with both Mark 1 (≤ -7.56 ± 6.52%) and Mark 2 (≤ 5.66 ± 1.41%) phantoms. For the UTHSCSA multimodality tumor measurement phantom, measured tumor volumes were highly correlated with NIST traceable design volumes for US (R2 = 1.000, p < 0.0001), CT (R2 = 0.9999, p < 0.0001) and MRI (R2 = 0.9998, p < 0.0001). CONCLUSIONS The UTHSCSA multimodality tumor measurement phantom described in this study can potentially be a useful quality assurance tool for verifying radiologic assessment of tumor size change during preclinical anti-cancer therapy testing with multiple imaging modalities.
منابع مشابه
A novel phantom design for brachytherapy quality assurance
Background: One major challenge in brachytherapy is to verify the accuracy of dose distributions calculated by the treatment planning system. In this project, a new phantom design has been introduced for quality assurance of dose distributions in gynocological (GYN) brachytherapy implants using EBT GafChromic film. Materials and Methods: This phantom has been designed and fabricated from 90 sla...
متن کاملDosimetric Study of an Indigenous and Heterogeneous Pelvic Phantom for Radiotherapy Quality Assurance
Introduction: In vitro dosimetric verification prior to patient treatment plays a key role in accurate and precision radiotherapy treatment delivery. Since the human body is a heterogeneous medium, the aim of this study was to design a heterogeneous pelvic phantom for radiotherapy quality assurance. Material and Methods: A pelvic phantom was ...
متن کاملQuality assurance prusedore in high dose rate bracythrapy
Introduction: Quality assurance is essential for obtaining the best achievable tumor control, avoiding unnecessary side effects, accuracy and safety during performing High Dose Rate (HDR) Brachytherapy (BT). Quality assurance is extremely important because HDR BT procedures are performed quickly, with high doses given in a short time period, with little opportunity for correcti...
متن کاملDesign and Fabrication of a Four-Dimensional Respiratory Phantom for Studying Tumor Movement in Radiotherapy with Magnetic Resonance Imaging
Introduction: In radiation therapy, determining the location of the tumor accurately during irradiation is one of the most important requirements. However, lung tumors are not fixed in a single location and move during irradiation due to respiratory motion. Due to limitations in assessing such movements, using a lung phantom can be useful and operational for their fast, easy an...
متن کاملA Specific Patient Quality Assurance (PSQA) procedure for a Co-60 source based High Dose Rate Brachytherapy
Introduction: In radiation therapy, accurate dose determination and precise dose delivery to the tumor are directly associated with better treatment outcomes in terms of higher tumor control and lower post radiation therapy complications. The current study aims the development and clinical application of the Patient Specific Quality Assurance (PSQA) procedures for nasopharyngea...
متن کامل