Friction in Total Hip Joint Prosthesis Measured In Vivo during Walking
نویسندگان
چکیده
Friction-induced moments and subsequent cup loosening can be the reason for total hip joint replacement failure. The aim of this study was to measure the in vivo contact forces and friction moments during walking. Instrumented hip implants with Al2O3 ceramic head and an XPE inlay were used. In vivo measurements were taken 3 months post operatively in 8 subjects. The coefficient of friction was calculated in 3D throughout the whole gait cycle, and average values of the friction-induced power dissipation in the joint were determined. On average, peak contact forces of 248% of the bodyweight and peak friction moments of 0.26% bodyweight times meter were determined. However, contact forces and friction moments varied greatly between individuals. The friction moment increased during the extension phase of the joint. The average coefficient of friction also increased during this period, from 0.04 (0.03 to 0.06) at contralateral toe off to 0.06 (0.04 to 0.08) at contralateral heel strike. During the flexion phase, the coefficient of friction increased further to 0.14 (0.09 to 0.23) at toe off. The average friction-induced power throughout the whole gait cycle was 2.3 W (1.4 W to 3.8 W). Although more parameters than only the synovia determine the friction, the wide ranges of friction coefficients and power dissipation indicate that the lubricating properties of synovia are individually very different. However, such differences may also exist in natural joints and may influence the progression of arthrosis. Furthermore, subjects with very high power dissipation may be at risk of thermally induced implant loosening. The large increase of the friction coefficient during each step could be caused by the synovia being squeezed out under load.
منابع مشابه
Postoperative Changes in In Vivo Measured Friction in Total Hip Joint Prosthesis during Walking
Loosening of the artificial cup and inlay is the most common reasons for total hip replacement failures. Polyethylene wear and aseptic loosening are frequent reasons. Furthermore, over the past few decades, the population of patients receiving total hip replacements has become younger and more active. Hence, a higher level of activity may include an increased risk of implant loosening as a resu...
متن کاملIn vivo measured joint friction in hip implants during walking after a short rest
INTRODUCTION It has been suspected that friction in hip implants is higher when walking is initiated after a resting period than during continuous movement. It cannot be excluded that such increased initial moments endanger the cup fixation in the acetabulum, overstress the taper connections in the implant or increase wear. To assess these risks, the contact forces, friction moments and frictio...
متن کاملMusculo-skeletal loading conditions at the hip during walking and stair climbing.
Musculo-skeletal loading plays an important role in the primary stability of joint replacements and in the biological processes involved in fracture healing. However, current knowledge of musculo-skeletal loading is still limited. In the past, a number of musculo-skeletal models have been developed to estimate loading conditions at the hip. So far, a cycle-to-cycle validation of predicted muscu...
متن کاملCementless Hip Arthroplasty in Southern Iran, Midterm Outcome and Comparison of Two Designs
Background: Cementless hip prosthesis was designed to provide biologic fixation, without the use of cement. The second generation components have shown more reliable bone ingrowths and survival rates. We are reporting a midterm result of two designs of cementless prosthesis in a unique culture with different social habits and expectations.Methods: 52 primary cementless total hip arthroplasty in...
متن کاملA Novel Engineering Spherical Bearing, with Potential Application for a Hip Implant
It is a research which could enter into contradiction with the current trend concerning the hip implants. It is known as sliding friction, which is characteristic of present artificial hip joints, is higher than in the case of rolling friction. The paper reports the studies of the functioning mode of a novel spherical bearing MoM (metal on metal) with rolling friction, with potential applicatio...