Comparison of anterior chamber depth measurement between Orbscan IIz and ultrasound biomicroscopy.
نویسندگان
چکیده
PURPOSE To compare the measurement of anterior chamber depth exclusive of corneal thickness using Orbscan IIz and ultrasound biomicroscopy (UBM) and evaluate the repeatability of each method. METHODS Three consecutive measurements of anterior chamber depth were prospectively performed using Orbscan IIz and Paradigm 50-MHz UBM in 40 eyes in 20 individuals. Mean values were compared using the paired t test. For 12 eyes in 6 individuals, anterior chamber depth measurements were performed 5 times to estimate the repeatability of each method by a coefficient of variation. Refractive errors were measured to correlate with anterior chamber depth. RESULTS The mean anterior chamber depth was 2.82 +/- 0.46 mm with the Orbscan IIz and 2.91 +/- 0.43 mm in UBM. This difference was statistically significant (P < .001), but not clinically meaningful. The coefficient of variation was 1.15% and 1.10% in Orbscan IIz and UBM, respectively. A negative correlation between anterior chamber depth and spherical equivalent refraction was noted within the range of -5.50 to +3.00 diopters. CONCLUSIONS The mean anterior chamber depth of Orbscan IIz was 0.087 mm less than that of UBM. Both methods were precise. Orbscan IIz seems to be a useful and more convenient method to measure anterior chamber depth for phakic intraocular lens implantation.
منابع مشابه
Agreement between Orbscan II, VuMAX UBM and Artemis-2 very-high frequency ultrasound scanner for measurement of anterior chamber depth
BACKGROUND The aim was to compare the anterior chamber depth (ACD) measurements taken with Orbscan II, ultrasound biomicroscopy (UBM) and the Artemis-2 VHF (very-high-frequency) ultrasound scanner in normal subjects. METHODS In this prospective study, one eye from each of 60 normal subjects was randomly selected. Three subjects dropped out of the study because they were apprehensive about the...
متن کاملComparison of anterior chamber depth measurements taken with the Pentacam, Orbscan IIz and IOLMaster in myopic and emmetropic eyes.
PURPOSE This study determined to assess the degree of agreement between anterior chamber depth (ACD) measurements obtained using three different devices and to analyse the relationship between ACD and spherical equivalent (SE) refraction. METHODS In this cross-sectional study, 42 eyes of 42 patients with a mean SE of - 4.69 +/- 4.61 D (range 0.00 D to - 14.88 D) were analysed. Measurements of...
متن کاملComparison of corneal measurements in keratoconic eyes using rotating Scheimpflug camera and scanning-slit topography.
AIM To compare the anterior segment measurements obtained by rotating Scheimpflug camera (Pentacam) and Scanning-slit topography (Orbscan IIz) in keratoconic eyes. METHODS A total of 121 patients, 71 males (58.7%) and 50 females (41.3%) (214 eyes) with the diagnosis of keratoconus (KC) were enrolled in this study. Following diagnosis of KC by slit-lamp biomicroscopic examination, central corn...
متن کاملComparison of anterior chamber depth in healthy myopic Thai eye using Orbscan and ultrasound.
OBJECTIVES To compare the anterior chamber depth (ACD) using ultrasonic and optical measurements. MATERIAL AND METHOD In this prospective study, ultrasound and optical (Orbscan) devices were employed to measure ACD in 42 eyes of 21 myopic volunteers. ACD values were compared using the paired-sample Student t test. The correlation of ACD values obtained from two groups was assessed by linear r...
متن کاملComparison of Central Corneal Thickness Measured through Pentacam, OCTCasia2, and Orbscan IIz in Normal Individuals
Purpose: To compare the central corneal thickness measured through Pentacam, OCTCasia2, and Orbscan IIz in normal individuals. Methods: In this cross sectional comparative study, the researcher investigated the eyes of 70 normal candidates (140 eyes) with the mean age of 31.42 ± 7.70, referring to Dr. Sedaghat eye clinic for refractive surgery. For each person, Central Corneal Thickness (CCT) w...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Journal of refractive surgery
دوره 23 5 شماره
صفحات -
تاریخ انتشار 2007