SU-E-T-24: Development and Implementation of an Automated Algorithm to Determine Radiation Isocenter, Radiation vs. Light Field Coincidence, and Analyze Strip Tests.

نویسندگان

  • D Hyer
  • C Mart
چکیده

PURPOSE The aim of this study was to develop a phantom and analysis software that could be used to quickly and accurately determine the location of radiation isocenter using the Electronic Portal Imaging Device (EPID). The phantom could then be used as a static reference point for performing other tests including: radiation vs. light field coincidence, MLC and Jaw strip tests, and Varian Optical Guidance Platform (OGP) calibration. METHODS The solution proposed uses a collimator setting of 10×10 cm to acquire EPID images of the new phantom constructed from LEGO® blocks. Images from a number of gantry and collimator angles are analyzed by the software to determine the position of the jaws and center of the phantom in each image. The distance between a chosen jaw and the phantom center is then compared to the same distance measured after a 180 degree collimator rotation to determine if the phantom is centered in the dimension being investigated. The accuracy of the algorithm's measurements were verified by independent measurement to be approximately equal to the detector's pitch. Light versus radiation field as well as MLC and Jaw strip tests are performed using measurements based on the phantom center once located at the radiation isocenter. RESULTS Reproducibility tests show that the algorithm's results were objectively repeatable. Additionally, the phantom and software are completely independent of linac vendor and this study presents results from two major linac manufacturers. An OGP calibration array was also integrated into the phantom to allow calibration of the OGP while the phantom is positioned at radiation isocenter to reduce setup uncertainty contained in the calibration. CONCLUSIONS This solution offers a quick, objective method to perform isocenter localization as well as laser alignment, OGP calibration, and other tests on a monthly basis.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

New Method of Quality Control Test for Light and Radiation Field Coincidence in Medical Linear Accelerators

Introduction: The evaluation of X-ray and light field coincidence in linear accelerators as a quality control test is often performed subjectively, involving the manual marking of films and their visual inspection following the irradiation. Therefore, the present study aimed to develop an objective method for the performance of this test leading to the increased levels...

متن کامل

New electrical device for field coincidence quality control test in linear accelerators

Testing of the coincidence of a linear accelerators X-ray field and the light field as a quality control test is often done in a subjective method, involving the manual marking of pieces of film and the visual inspection of the film after irradiation. The purpose of this study was to develop an objective method for performing this test, while also increasing the accuracy, precision and the spee...

متن کامل

SU-E-T-184: LINAC Commissioning Measurements Utilizing the Cylindrical Arc Check Phantom.

PURPOSE New devices for IMRT and VMAT QA are continually becoming available. Despite the availability of these devices they are underutilized for routine mechanical QA. In this work we take advantage of the cylindrical geometry of the Arc Check phantom to perform tests that are an integral part of commissioning and routine QA. METHODS The Arc Check is a cylindrical phantom with a diameter of ...

متن کامل

Development of a prototype stereotactic collimation assembly for Neptun 10 PC linac

Background: A set of hardware stereotactic radiosurgery is designed and constructed. The scope of this article is to describe how we have designed and developed stereotactic collimation assembly for a 9 MV Neptun 10 PC linac. Materials and methods: One collimator holder and six collimators were made, with their radiation field diameter between 12.5 mm to 25 mm at isocenter level. Beam data, suc...

متن کامل

SU-F-T-493: An Investigation Into the Feasibility of Using PipsPro Software with Film for Linac QA.

PURPOSE To determine the feasibility of using radiochromic and radiographic film with Pipspro software for quality assurance of linear accelerators with no on-board imaging. METHODS The linear accelerator being used is a Varian Clinac 21EX. All IGRT is performed using the BrainLab ExacTrac system. Because of the lack of on board imaging, certain monthly and annual TG-142 quality assurance tes...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Medical physics

دوره 39 6Part9  شماره 

صفحات  -

تاریخ انتشار 2012