Factors Affecting the Accuracy of Intraocular Lens Power Calculation with Lenstar

نویسندگان

  • Tae Hee Lee
  • Mi Sun Sung
  • Lian Cui
  • Ying Li
  • Kyung Chul Yoon
چکیده

This retrospective study was performed to compare refractive outcomes measured by conventional methods and by use of the Lenstar biometer and to investigate the factors affecting intraocular lens (IOL) power calculation with Lenstar with and without IOL-constant optimization. The study included 100 eyes of 86 patients who underwent cataract surgery. Corneal curvature was measured with a manual keratometer (MK), automated keratometer (AK), and the Lenstar biometer, and axial length (AL) was measured by A-scan and Lenstar. Mean numerical error (MNE) and mean absolute error (MAE) were compared between AK and MK with A-scan, and Lenstar with and without optimization. Factors affecting the accuracy of the IOL power calculation by use of Lenstar with and without optimization were analyzed. No significant differences were observed in the MNE or MAE among the devices. The proportion of MAE within 0.5 D was higher for Lenstar with optimization (62.7%) than without optimization (46.2%). The proportion of MAE within 0.5 D was 62% and 58% for MK and AK with A-scan, respectively. Without optimization, the MAE was smaller in eyes with ALs between 23 mm and 25 mm (p=0.03), whereas it was smaller at higher corneal powers when the IOL constant was optimized (>44 D, p=0.03). The IOL power calculations showed no significant differences among the devices, but the results of MAE within 0.5 D by use of Lenstar without optimization were worse than those of conventional methods. The AL influenced the accuracy of refractive outcomes determined by using Lenstar without optimization, and corneal curvature was shown to affect the accuracy of refractive measurements using Lenstar with optimization.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

مقایسه‌ی کراتومتری پنتاکم وکراتومتری IOL Master در محاسبه‌ی لنز داخل چشمی افراد با قرنیه‌ی طبیعی

Background and Objective: Proper method and machine for corneal evaluation is an important factor in many anterior segment interventions. This study was performed to compare the mean keratometry (K) readings obtained with an IOL Master and Pentacam for intraocular lens power calculation in normal subjects with no history of refractive surgery. Materials and Methods: Mean K values were obtained ...

متن کامل

Accuracy of Various Intraocular Lens Power Calculation Formulas in Steep Corneas

PURPOSE To compare the accuracy of four different intraocular lens (IOL) power calculation formulas for eyes with mean keratometry values greater than 46 diopters (D). METHODS Forty five eyes from 45 patients who were candidates for senile cataract surgery with mean keratometry values greater than 46 D were included. Calculation of the IOL power was performed by the Lenstar. The implanted IOL...

متن کامل

مقایسه دو روش بیومتری التراسوند (Ultrasound) و لنستار (Lenstar) در تعیین طول محور چشم و میزان عیب انکساری پس از عمل جراحی کاتاراکت

Background and Objective: Ocular biometry refers to the measurement of the axial length of the eye and thickness of the intraocular structures. This process is routinely performed for all patients before cataract surgery. The accuracy of the biometric data is directly associated with the refractive status after cataract surgery. Currently, two methods of biometry, namely ultrasound and optical ...

متن کامل

Intraocular lens calculation for aspheric intraocular lenses.

PURPOSE To evaluate the possible benefits of biometry and ray-tracing intraocular lens (IOL) calculation for aspheric aberration-correcting IOLs. SETTING Private eye clinic in Germany. DESIGN Retrospective consecutive case series. METHODS Eyes with 3 different aberration-correcting IOLs were reviewed. Before surgery, the axial length, corneal thickness, anterior chamber depth, crystalline...

متن کامل

Visual and Refractive Results of Cataract Surgery by Phacoemulsification Method in Patients with Keratoconus

Purpose: To examine the visual and refractive results of cataract surgery by phacoemulsification method in patients with keratoconus. Methods: In this retrospective study, the files of patients with keratoconus, who underwent phacoemulsification between 2008 and 2017, were evaluated. The intraocular lens power calculation formula was SRK/T. Post-operative visual and refractive results were anal...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 51  شماره 

صفحات  -

تاریخ انتشار 2015