Biomechanical Analysis of a Pedicle Screw-Rod System with a Novel Cross-Link Configuration

نویسندگان

  • Yasuhiro Nakajima
  • Masahito Hara
  • Daisuke Umebayashi
  • Shoichi Haimoto
  • Yu Yamamoto
  • Yusuke Nishimura
  • Toshihiko Wakabayashi
چکیده

STUDY DESIGN The strength effects of a pedicle screw-rod system supplemented with a novel cross-link configuration were biomechanically evaluated in porcine spines. PURPOSE To assess the biomechanical differences between a conventional cross-link pedicle screw-rod system versus a novel cross-link instrumentation, and to determine the effect of the cross-links. OVERVIEW OF LITERATURE Transverse cross-link systems affect torsional rigidity, but are thought to have little impact on the sagittal motion of spinal constructs. We tested the strength effects in pullout and flexion-compression tests of novel cross-link pedicle screw constructs using porcine thoracic and lumbar vertebrae. METHODS Five matched thoracic and lumbar vertebral segments from 15 porcine spines were instrumented with 5.0-mm pedicle screws, which were then connected with 6.0-mm rods after partial corpectomy in the middle vertebral body. The forces required for construct failure in pullout and flexion-compression tests were examined in a randomized manner for three different cross-link configurations: un-cross-link control, conventional cross-link, and cross-link passing through the base of the spinous process. Statistical comparisons of strength data were analyzed using Student's t-tests. RESULTS The spinous process group required a significantly greater pullout force for construct failure than the control group (p=0.036). No difference was found between the control and cross-link groups, or the cross-link and spinous process groups in pullout testing. In flexion-compression testing, the spinous processes group required significantly greater forces for construct failure than the control and cross-link groups (p<0.001 and p=0.003, respectively). However, there was no difference between the control and cross-link groups. CONCLUSIONS A novel cross-link configuration that features cross-link devices passing through the base of the spinous processes increased the mechanical resistance in pullout and flexion-compression testing compared to un-cross-link constructs. This configuration provided more resistance to middle-column damage under flexion-compression testing than conventional cross-link configuration.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Biomechanical Stability of a Cross-Rod Connection with a Pedicle Screw System

BACKGROUND Surgery with pedicle screw instrumentation does not provide sufficient torsional stability. This leads to pseudoarthrosis, loosening of the pedicle screws, and, ultimately, implant failure. MATERIAL AND METHODS Functional spinal units from 18 deer were evaluated using a 6-axis material testing machine. As specimen models, we prepared an intact model, a damaged model, a cross-rod mode...

متن کامل

Biomechanical Evaluation of a Novel Posterior Integrated Clamp That Attaches to an Existing Posterior Instrumentation for Use in Thoracolumbar Revision

STUDY DESIGN An in vitro biomechanical study. PURPOSE To evaluate the biomechanics of a novel posterior integrated clamp (IC) that extends on an already implanted construct in comparison to single long continuous bilateral pedicle screw (BPS) and rod stabilization system. OVERVIEW OF LITERATURE Revision surgery in the thoracolumbar spine often necessitates further instrumentation following ...

متن کامل

A biomechanical comparison of a novel thoracic screw fixation method: transarticular screw fixation vs traditional pedicle screw fixation.

BACKGROUND Transarticular screw fixation is used in the upper cervical and lumbar spine to achieve posterior spinal stability, and its biomechanical performance has been proven to be similar to that of pedicle screw fixation. However, few studies have reported the use of transarticular screw fixation in the upper thoracic spine. OBJECTIVE To biomechanically compare transarticular screws with ...

متن کامل

Comparative Analysis of Effect of Density, Insertion Angle and Reinsertion on Pull-Out Strength of Single and Two Pedicle Screw Constructs Using Synthetic Bone Model

STUDY DESIGN Biomechanical study. PURPOSE To determine the effect of density, insertion angle and reinsertion on pull-out strength of pedicle screw in single and two screw-rod configurations. OVERVIEW OF LITERATURE Pedicle screw pull-out studies have involved single screw construct, whereas two screws and rod constructs are always used in spine fusions. Extrapolation of results using the si...

متن کامل

Novel Pedicle Screw and Plate System Provides Superior Stability in Unilateral Fixation for Minimally Invasive Transforaminal Lumbar Interbody Fusion: An In Vitro Biomechanical Study

PURPOSE This study aims to compare the biomechanical properties of the novel pedicle screw and plate system with the traditional rod system in asymmetrical posterior stabilization for minimally invasive transforaminal lumbar interbody fusion (MI-TLIF). We compared the immediate stabilizing effects of fusion segment and the strain distribution on the vertebral body. METHODS Seven fresh calf lu...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 10  شماره 

صفحات  -

تاریخ انتشار 2016