Aggressive pacemaker twiddler's syndrome. Dislodgement of an active fixation ventricular pacing electrode.

نویسندگان

  • R B Lal
  • R D Avery
چکیده

Pacemaker twiddler's syndrome is the phenomenon whereby a pacemaker patient may actively dislodge a pacing electrode by manipulating, often unconsciously, the permanent implanted pacemaker. This unusual case is remarkable in several respects: surgical precautions, originally taken to secure the pulse generator against migration, proved ineffective; the displaced lead featured an active fixation "screw-in" electrode; and the reported dislodgement occurred not in the early postimplant stage, but fully three months later. It can be concluded that even with advanced lead technology and operative precaution, electrode dislodgement cannot be absolutely prevented. Awareness of this phenomenon, especially by pacemaker follow-up care personnel, could lessen its occurrence.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Pacemaker Twiddler's Syndrome

A 76-yr-old man with unipolar ventricular pacemaker, previously implanted in the right ventricular apex, presented with syncopal attack and twitching of the right pectoralis muscle. He finally was diagnosed as having Twiddler's syndrome. The patient had a pulse rate of 36/min and cannon waves in the neck veins. Chest radiography showed a dislodged pacing lead completely wrapped around the pulse...

متن کامل

Twiddler's syndrome: a rare cause of pacemaker failure.

160 Twiddler's syndrome was first described by Bayliss et al. in 1968 [1]. This condition is a rare cause of pacemaker failure, resulting from the dislodging of pacing electrodes by manipulation, often unconsciously, of the implanted pacemaker [2]. This report describes a rare condition of pacemaker failure in a patient with twiddler's syndrome resulting from unintentional rotation of the pacem...

متن کامل

Ratchet-traction effect: an underdiagnosed mechanism of pacing lead dislodgement.

Pacing lead dislodgement contributes substantially, to the list of causes of early pacemaker failure. Reel's syndrome is a rare cause of pacemaker failure, resulting from the dislodging of pacing electrodes by manipulation of the pulse generator by the patient. We describe here an intriguing case of pacemaker lead dislodgement by a novel and frequently underdiagnosed mechanism that mimicked the...

متن کامل

Comparison of mid-term clinical experience with steroid-eluting active and passive fixation ventricular electrodes in children.

Although active fixation ventricular leads seem to have advantages over passive fixation leads, this study compares the follow-up results of active and passive fixation leads in children. We evaluated the implantation and follow-up data of 41 children with active (Accufix II DEC, group 1) (n = 20) or passive (Membrane E, group 2) (n = 21) fixation, steroid-eluting ventricular leads. All but one...

متن کامل

Iatrogenic Twiddler's Syndrome: Case Report and Proposed Experimental Model

A 74-year old woman, with history of systemic hypertension, dyslipemia, obesity and sick sinus syndrome (tachy-brady syndrome) underwent single chamber pacemaker implantation (Sensia SR, Capsure 5054 lead, Medtronic Inc, Minneapolis, MN, USA) via left subclavian vein puncture. The parameters at the implant were Rwave 12 mV, pacing threshold 0,4 V at 0,5ms, impedance 1200 Ω. During the procedure...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Chest

دوره 97 3  شماره 

صفحات  -

تاریخ انتشار 1990