CT fluoroscopy shielding: decreases in scattered radiation for the patient and operator.
نویسندگان
چکیده
PURPOSE High-radiation exposure occurs during computed tomographic (CT) fluoroscopy. Patient and operator doses during thoracic and abdominal interventional procedures were studied in the present experiment, and a novel shielding device to reduce exposure to the patient and operator was evaluated. MATERIALS AND METHODS With a 16-slice CT scanner in CT fluoroscopy mode (120 kVp, 30 mA), surface dosimetry was performed on adult and pediatric phantoms. The shielding was composed of tungsten antimony in the form of a lightweight polymer sheet. Doses to the patient were measured with and without shielding for thoracic and abdominal procedures. Doses to the operator were recorded with and without phantom, gantry, and table shielding in place. Double-layer lead-free gloves were used by the operator during the procedures. RESULTS Tungsten antimony shielding adjacent to the scan plane resulted in a maximum dose reduction of 92.3% to the patient. Maximum 85.6%, 93.3%, and 85.1% dose reductions were observed for the operator's torso, gonads, and hands, respectively. The use of double-layer lead-free gloves resulted in a maximum radiation dose reduction of 97%. CONCLUSIONS Methods to reduce exposure during CT fluoroscopy are effective and should be searched for. Significant reduction in radiation doses to the patient and operator can be accomplished with tungsten antimony shielding.
منابع مشابه
Estimation of the effective dose to the radiologists during fluoroscopy or angiography of abdominal viscera
ABSTRACTBackground: as fluoroscopy or different angiographies are usually evaluated using TL dosimeter. Thismethod is a time-consuming procedure therefore, radiologists are usually worried and unawareof their exact radiation dose received during each fluoroscopy or angiography. In this study, anew trend for quick estimation of effective dose based on measuring air-absorbed dose of thescattered ...
متن کاملEvaluation of Radiation Exposure to Staff and Environment Dose from [18F]-FDG in PET/CT and Cyclotron Center using Thermoluminescent Dosimetry
Background: PET/CT imaging using [18F]-FDG is utilized in clinical oncology for tumor detecting, staging and responding to therapy procedures. Essential consideration must be taken for radiation staff due to high gamma radiation in PET/CT and cyclotron center. The aim of this study was to assess the staff exposure regarding whole body and organ dose and to evaluate environment dose in PET/CT an...
متن کاملA novel radiation protection device based on tungsten functional paper for application in interventional radiology
Tungsten functional paper (TFP), which contains 80% tungsten by weight, has radiation-shielding properties. We investigated the use of TFP for the protection of operators during interventional or therapeutic angiography. The air kerma rate of scattered radiation from a simulated patient was measured, with and without TFP, using a water-equivalent phantom and fixed C-arm fluoroscopy. Measurement...
متن کاملComparing strategies for operator eye protection in the interventional radiology suite.
PURPOSE To evaluate the impact of common radiation-shielding strategies, used alone and in combination, on scattered dose to the fluoroscopy operator's eye. MATERIALS AND METHODS With an operator phantom positioned at the groin, upper abdomen, and neck, posteroanterior low-dose fluoroscopy was performed at the phantom patient's upper abdomen. Operator lens radiation dose rate was recorded wit...
متن کاملApplication of Different methods for Reducing Radiation Dose to Breast during MDCT
The increased use of computed tomography (CT) and its high radiation dose have led to great concerns about its potential for radiation induced cancer risks. Breast is a radiosensitive tissue based on tissue weighting factors assigned by the International Commission on Radiological Protection (ICRP). Moreover, the dose is maximal on the surface of the patient. Therefore, strategies should be tak...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Journal of vascular and interventional radiology : JVIR
دوره 17 12 شماره
صفحات -
تاریخ انتشار 2006