Performance characteristics of a polyclonal enzyme immunoassay for the quantitation of Histoplasma antigen in human urine samples.
نویسندگان
چکیده
An antigen detection assay was prepared with rabbit anti- Histoplasma antibodies to detect and quantitate Histoplasma capsulatum antigen in urine samples. By using a 4-parameter curve fit, the assay calibration ranges from 2 to 1,000 enzyme immunoassay (EIA) units. We compared results for 99 urine samples with those of a reference laboratory, half of which tested positive or equivocal by that reference laboratory. Performance characteristics were further defined by studying the assay linearity, precision, percentage of positive agreement, and percentage of negative agreement. An acceptable correlation with the reference laboratory (R2=0.7174) was obtained with results ranging from less than 2 (negative samples) to 132 EIA units by the new method. Compared with the reference laboratory, the percentage of agreement for positive samples, excluding equivocal samples, was 92% and for negative samples was 98%. Crossreactivity occurred with culture filtrates of Paracoccidioides brasiliensis, Coccidioides immitis, and Blastomyces dermatiditis. No cross-reactivity was observed with Candida albicans or Aspergillus fumigatus culture filtrates. The current EIA for the detection and quantitation of H capsulatum antigen in urine specimens is a valid assay that agrees well with results from an established reference laboratory.
منابع مشابه
Evaluation of an enzyme immunoassay for detection of Histoplasma capsulatum antigen from urine specimens.
Detection of Histoplasma capsulatum urinary antigen (UAg) is important for the initial diagnosis of infection and for monitoring of patient responses to antifungal therapy. This study evaluated an analyte-specific reagent (ASR) enzyme immunoassay (EIA) for the detection of H. capsulatum UAg from Immuno Mycologics, Inc. (IMMY) (Norman, OK) in comparison with routine testing with the MiraVista (M...
متن کاملDetection of histoplasma antigen by a quantitative enzyme immunoassay.
The second-generation Histoplasma antigen immunoassay is semiquantitative, expressing results as a comparison to a negative control, which requires repeat testing of the prior specimen with the current specimen to accurately determine a change in antigen. Reporting results in this manner often is confusing to the ordering physician and laboratory. Development of a quantitative assay could impro...
متن کاملEvaluation of commercially available reagents for diagnosis of histoplasmosis infection in immunocompromised patients.
Urinary histoplasma antigen measurement can be useful for diagnosing systemic histoplasmosis and for monitoring treatment response, especially in immunocompromised patients. However, testing has traditionally been limited to specialized reference laboratories, as immunoassay reagents for the antigen were not widely available. Recently, a polyclonal-antibody-based in vitro diagnostic (IVD) kit f...
متن کاملAntigenemia without antigenuria in a cat with histoplasmosis
CASE SUMMARY Based on demonstration of the yeast phase of Histoplasma capsulatum on fine-needle aspirate cytology of the kidney, a 5-year-old cat was diagnosed with histoplasmosis. Urine and serum were tested for antigen via a Histoplasma antigen enzyme immunoassay. At the time of diagnosis, and on multiple occasions during antifungal treatment, antigenemia was detected without antigenuria. The...
متن کاملCryptococcal glucoxylomannan does not exhibit cross-reactivity in the MVista Histoplasma antigen enzyme immunoassay.
The potential for cross-reaction between Cryptococcus neoformans and Histoplasma capsulatum in antigen assays was evaluated. We tested patient samples, spleens from infected mice, and purified polysaccharides in the MVista Histoplasma antigen enzyme immunoassay and cryptococcal antigen latex agglutination system for cross-reactivity, and none was observed.
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- American journal of clinical pathology
دوره 128 1 شماره
صفحات -
تاریخ انتشار 2007