A new use for the amplatzer duct occluder device.

نویسندگان

  • Carlos A C Pedra
  • Luciano N Sousa
  • Carlo B Pilla
  • Valmir F Fontes
چکیده

We report a case in which the Amplatzer device for percutaneous occlusion of ductus arteriosus was successfully used for occluding a large systemic-pulmonary collateral vessel in a patient who had previously undergone surgery for correction of pulmonary atresia and ventricular septal defect (Rastelli technique), and was awaiting the change of a cardiac tube. In the first attempt, the device embolized to the distal pulmonary bed and, after being rescued with a Bitome, it was appropriately repositioned with no complications and with total occlusion of the vessel.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Transcatheter closure of patent arterial duct with amplatzer duct occluder II additional sizes in children.

Amplatzer duct occluder II additional sizes is a new generation of device for arterial duct closure, appropriate for infants and older children. We report the successful transcatheter closure of the arterial duct in an infant and two older children with the above-mentioned device.

متن کامل

Percutaneous Forceps Retrieval of an Embolized Amplatzer Duct Occluder

Transcatheter closure of patent ductus arteriosus with the Amplatzer duct occluder shows excellent results. However, there is a risk of device embolization after deployment. We report the case of a 33-year-old man in whom an embolized Amplatzer device was retrieved from the right pulmonary artery by a percutaneous method. We also review this uncommon severe complication and offer technical tips...

متن کامل

Transcatheter closure of left coronary cameral fistula with Amplatzer duct occluder II.

The surgical and transcatheter coil closure of coronary arterial fistulas are described in the literature. We report our experience with the successful transcatheter closure of a coronary arterial fistula, arising from the left coronary artery and draining into the right ventricle, with the new Amplatzer duct occluder II device.

متن کامل

Significant persistent ductus arteriosus in infants less or equal to 6 kg: percutaneous closure or surgery?

BACKGROUND Percutaneous closure of large persistent ductus arteriosus using the Amplatzer duct occluder is an alternative to surgery. However, this device is not recommended in infants weighing less than 6 kg. AIM To evaluate the safety and effectiveness of this procedure in low-body-weight infants. METHODS We reviewed retrospectively data for infants weighing less or equal to 6 kg who unde...

متن کامل

Transcatheter closure of persistent ductus arteriosus in infants using the Amplatzer duct occluder.

AIM To evaluate whether transcatheter closure with the Amplatzer duct occluder offers an alternative to surgical treatment in infants with a persistent ductus arteriosus. METHODS 12 patients under 1 year of age (age 1-11 months, body weight 2.6-8.7 kg) with clinical and echocardiographic findings of a significant duct were considered for transcatheter closure with the Amplatzer occluder. The ...

متن کامل

Safety and efficacy of Amplatzer duct occluder for percutaneous closure of ventricular septal defects with tunnel shape aneurysm: Medium term follow up

Objectives: Different devices including Amplatzer duct occluder has been used for percutaneous closure of ventricular septal defects. This study reports our medium term follow up of perimembranous and muscular ventricular septal defects with tunnel shape aneurysm closure using the Amplatzer duct occluder. Materials and Methods: From May 2006-December 2012, we used Amplatzer duct occluder in sev...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Arquivos brasileiros de cardiologia

دوره 78 3  شماره 

صفحات  -

تاریخ انتشار 2002