Dose calibrator linearity test: 99mTc versus 18F radioisotopes*

نویسندگان

  • José Willegaignon
  • Marcelo Tatit Sapienza
  • George Barberio Coura-Filho
  • Alexandre Teles Garcez
  • Carlos Eduardo Gonzalez Ribeiro Alves
  • Marissa Anabel Rivera Cardona
  • Ricardo Fraga Gutterres
  • Carlos Alberto Buchpiguel
چکیده

OBJECTIVE The present study was aimed at evaluating the viability of replacing (18)F with (99m)Tc in dose calibrator linearity testing. MATERIALS AND METHODS The test was performed with sources of (99m)Tc (62 GBq) and (18)F (12 GBq) whose activities were measured up to values lower than 1 MBq. Ratios and deviations between experimental and theoretical (99m)Tc and (18)F sources activities were calculated and subsequently compared. RESULTS Mean deviations between experimental and theoretical (99m)Tc and (18)F sources activities were 0.56 (± 1.79)% and 0.92 (± 1.19)%, respectively. The mean ratio between activities indicated by the device for the (99m)Tc source as measured with the equipment pre-calibrated to measure (99m)Tc and (18)F was 3.42 (± 0.06), and for the (18)F source this ratio was 3.39 (± 0.05), values considered constant over the measurement time. CONCLUSION The results of the linearity test using (99m)Tc were compatible with those obtained with the (18)F source, indicating the viability of utilizing both radioisotopes in dose calibrator linearity testing. Such information in association with the high potential of radiation exposure and costs involved in (18)F acquisition suggest (99m)Tc as the element of choice to perform dose calibrator linearity tests in centers that use (18)F, without any detriment to the procedure as well as to the quality of the nuclear medicine service.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

A Recommendation for Revised Dose Calibrator Measurement Procedures for 89Zr and 124I

Because of their chemical properties and multiday half lives, iodine-124 and zirconium-89 are being used in a growing number of PET imaging studies. Some aspects of their quantitation, however, still need attention. For (89)Zr the PET images should, in principle, be as quantitatively accurate as similarly reconstructed 18F measurements. We found, however, that images of a 20 cm well calibration...

متن کامل

18F Protection Issues: Human and -Camera Considerations

Objective: Two main issues in protecting radiation workers and the general public from 18F radiation—distance from and lead shielding for an 18F source—were investigated. We also examined the effect of an 18F source on the counting rate of a neighboring -camera. Methods: The dose rates of an 18F vial and a water-filled cylinder were measured using an ionization chamber at different distances wi...

متن کامل

Dose Calibrator Linearity Testing Using an Improved Attenuator System

Nuclear Regulatory Commission (NRC) guidelines state that dose calibrators must be tested for linearity upon installation and quarterly thereafter over a range of activities from 10 ~J,Ci (370 kBq) to the highest dose administered to the patient (300 mCi [11.1 GBq] ofiodine-131 for therapy at our institution). The simplest method of linearity testing utilizes commercial attenuator kits. These k...

متن کامل

Quality Control Testing of Dose Calibrators

We have reviewed quality control testing of dose calibrators and have analyzed the potential for error. The uncertainty in half-lives of reference sources can lead to inaccurate assessment of the accuracy and constancy in dose calibrator performance. Molybdenum-99 contamination of '"Tc used as a reference source can appear as a nonlinear response of the dose calibrator. The useful life of a ref...

متن کامل

Calibration Test of PET Scanners in a Multi-Centre Clinical Trial on Breast Cancer Therapy Monitoring Using 18F-FLT

UNLABELLED A multi-centre trial using PET requires the analysis of images acquired on different systems We designed a multi-centre trial to estimate the value of 18F-FLT-PET to predict response to neoadjuvant chemotherapy in patients with newly diagnosed breast cancer. A calibration check of each PET-CT and of its peripheral devices was performed to evaluate the reliability of the results. MA...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 48  شماره 

صفحات  -

تاریخ انتشار 2015