The desaturation response time of finger pulse oximeters during mild hypothermia.
نویسندگان
چکیده
Pulse oximeters may delay displaying the correct oxygen saturation during the onset of hypoxia. We investigated the desaturation response times of pulse oximeter sensors (forehead, ear and finger) during vasoconstriction due to mild hypothermia and vasodilation caused by glyceryl trinitrate. Ten healthy male volunteers were given three hypoxic challenges of 3 min duration under differing experimental conditions. Mild hypothermia increased the mean response time of finger oximeters from 130 to 215 s. Glyceryl trinitrate partly offset this effect by reducing the response time from 215 to 187 s. In contrast, the response times of the forehead and ear oximeters were unaffected by mild hypothermia, but the difference between head and finger oximeters was highly significant (p < 0.0001). The results suggest that the head oximeters provide a better monitoring site for pulse oximeters during mild hypothermia.
منابع مشابه
The effects of motion artifact and low perfusion on the performance of a new generation of pulse oximeters in volunteers undergoing hypoxemia.
INTRODUCTION Motion artifact and low perfusion often lead to faulty or absent pulse oximetry readings in clinical practice. OBJECTIVE Determine the impact of motion artifact and low perfusion on newly introduced pulse oximetry technologies during hypoxemic episodes in healthy volunteers. METHODS Five different pulse oximeters from 4 manufacturers (the Datex Ohmeda 3900P; the Agilent; the Ne...
متن کاملMeta-analysis of arterial oxygen saturation monitoring by pulse oximetry in adults.
OBJECTIVE The purposes of the study were to: (1) describe the aggregate strength of the relationship of arterial oxygen saturation as measured by pulse oximetry with the standard of arterial blood gas analysis as measured by co-oximetry, (2) examine how various factors affect this relationship, and (3) describe an aggregate estimate of the bias and precision between oxygen saturation as measure...
متن کاملArterial Blood Oxygen Saturation during Blood Pressure Cuff-induced Hypoperfusion
Pulse oximetry has been one of the most significant technological advances in clinical monitoring in the last two decades. Pulse oximetry is a non-invasive photometric technique that provides information about the arterial blood oxygen saturation (SpO2) and heart rate, and has widespread clinical applications. When peripheral perfusion is poor, as in states of hypovolaemia, hypothermia and vaso...
متن کاملEffects of skin pigmentation on pulse oximeter accuracy at low saturation.
BACKGROUND It is uncertain whether skin pigmentation affects pulse oximeter accuracy at low HbO2 saturation. METHODS The accuracy of finger pulse oximeters during stable, plateau levels of arterial oxygen saturation (Sao2) between 60 and 100% were evaluated in 11 subjects with darkly pigmented skin and in 10 with light skin pigmentation. Oximeters tested were the Nellcor N-595 with the OxiMax...
متن کاملSleep desaturation: comparison of two oximeters.
Oxygen saturation is measured by pulse oximetry during sleep studies. Body movements and peripheral vasoconstriction related to respiratory events may interfere with measurements by conventional oximeters. Our objective was to compare the detection rate of sleep desaturations by two oximeters, one of which used new motion-resistant technology. We studied 34 children (median age, 13 years; range...