SCIENTIFIC REPORT Intraobserver and interobserver reproducibility in the evaluation of ultrasonic pachymetry measurements of central corneal thickness
نویسندگان
چکیده
Purpose: To assess reproducibility of central corneal thickness (CCT) measurement by means of ultrasonic pachymetry. Methods: Fifty one volunteers underwent three sessions of CCT measurements, each consisting of three CCT measurements, performed by each of three different observers. Intraand interobserver reproducibility was calculated by means of intraclass correlation coefficient (ICC). The expected range of variability between two independent evaluations was calculated using scatter plots of each test-retest difference against their mean. The standard deviation of the mean differences in the test-retest scores was used to describe the differences in the score spread. Results: The ICC ranges of the intraand interobserver evaluations were 0.95–0.97 and 0.89–0.95 respectively; the expected variability was (¡1% and (¡ 2% respectively (95% confidence interval). Conclusions: The measurement of CCT by means of ultrasonic pachymetry is highly reproducible.
منابع مشابه
Intraobserver and interobserver reproducibility in the evaluation of ultrasonic pachymetry measurements of central corneal thickness.
PURPOSE To assess reproducibility of central corneal thickness (CCT) measurement by means of ultrasonic pachymetry. METHODS Fifty one volunteers underwent three sessions of CCT measurements, each consisting of three CCT measurements, performed by each of three different observers. Intra- and interobserver reproducibility was calculated by means of intraclass correlation coefficient (ICC). The...
متن کاملComparison of central corneal thickness measurements by rotating Scheimpflug camera, ultrasonic pachymetry, and scanning-slit corneal topography.
PURPOSE To compare central corneal thickness measurements and their reproducibility when taken by a rotating Scheimpflug camera, ultrasonic pachymetry, and scanning-slit corneal topography/pachymetry. DESIGN Experimental study. PARTICIPANTS Seventy-four eyes of 64 subjects without ocular abnormalities other than cataract. METHODS Corneal thickness measurements were compared among the 3 me...
متن کاملSCIENTIFIC REPORT Comparison of corneal thickness measurements using Orbscan II, non-contact specular microscopy, and ultrasonic pachymetry in eyes after laser in situ keratomileusis
Aims: To compare central corneal thickness measurements of three pachymetry devices in eyes after laser in situ keratomileusis (LASIK). Methods: Central corneal thickness was measured in 203 eyes after myopic LASIK. Orbscan II scanning slit topography (Bausch & Lomb), SP-2000P non-contact specular microscopy (Topcon), and ultrasonic pachymetry (Tomey) were used in this sequence. Results: Three ...
متن کاملReproducibility and clinical evaluation of rebound tonometry.
PURPOSE To establish the reproducibility of a rebound tonometer in humans and the effect of corneal thickness on measurements, comparing it with Goldmann applanation tonometer. METHODS In a first study designed to examine the reliability of the RBT, three experienced ophthalmologists undertook three consecutive intraocular pressure (IOP) measurements in 12 eyes of 12 normal subjects. A cross-...
متن کاملRepeatability and reproducibility of pachymetric mapping with Visante anterior segment-optical coherence tomography.
PURPOSE To determine the repeatability and reproducibility of central and peripheral corneal pachymetry mapping with anterior segment-optical coherence tomography (AS-OCT). METHODS An observational cross-sectional study involving two groups: 27 healthy eyes and 20 eyes with keratoconus. Each subject underwent scanning sessions with AS-OCT to determine intraobserver repeatability, interobserve...
متن کامل