[Comparison between Goldmann, Icare Pro and Corvis ST tonometry].

نویسندگان

  • P Bañeros-Rojas
  • J M Martinez de la Casa
  • P Arribas-Pardo
  • C Berrozpe-Villabona
  • P Toro-Utrera
  • J García-Feijoó
چکیده

PURPOSE To compare intraocular pressure (IOP) between the new non-contact tonometer Corvis ST (CST), the Goldmann applanation tonometry (GAT) and Icare Pro rebound tonometer (PRO). METHODS A total of 178 eyes of 178 healthy subjects were selected for the study. Measurements of IOP were made in a random order with GAT, PRO and CST. Central corneal thickness (CCT) was determined by ultrasound pachymetry. The mean of three valid measurements of each variable was used in the statistical analysis. The relationship between the tonometers was established using Bland-Altman plots. RESULTS Mean IOP was 15.5 ± 2.8 mmHg for GAT, 15.4 ± 2.8 mmHg for CST, and 14.6 ± 2.3 mmHg for PRO. The mean differences between pairs of tonometers were: GAT-PRO=0.9 ± 1.7 mmHg (P<.001), GAT-CST: 0.1 ± 2.2 (P=.398), and PRO-CST: -0.8 ±-0.7 mmHg, p<0.001. A positive relationship was detected between CCT and the three tonometers: GAT: r=0.325, P<.001; PRO: r = 0.385, P<.001, and CST: r = 0.428; P<.001. CONCLUSIONS The differences found between PRO and GAT were significantly higher than those found between CST and GAT, which showed non-significant differences. The measurements of the three tonometers were affected by the CCT.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Comparison of three intraocular pressure measurement methods including biomechanical properties of the cornea.

PURPOSE The aim of this study was to show the usefulness of three methods for measuring IOP: Goldmann applanation tonometry, rebound tonometry, and Ultra-High-Speed Scheimpflug technology. METHODS The examined group consisted of 96 patients (192 eyes), including 63 women and 33 men with a mean age of 59.3 ± 19.9 years. Together, 152 healthy eyes and 40 eyes with different pathologies were exa...

متن کامل

Erratum: A cross-sectional study to compare intraocular pressure measurement by sequential use of Goldman applanation tonometry, dynamic contour tonometry, ocular response analyzer, and Corvis ST

OBJECTIVE To study the correlation and effect of sequential measurement of intraocular pressure (IOP) with Goldmann applanation tonometer (GAT), ocular response analyzer (ORA), dynamic contour tonometer (DCT), and Corvis ST. SETTING AND DESIGN Observational cross-sectional series from the comprehensive clinic of a tertiary eye care center seen during December 2012. METHODS One hundred and t...

متن کامل

A Comparative Study of Rebound Tonometry With Tonopen and Goldmann Applanation Tonometry Following Vitreoretinal Surgery.

PURPOSE To investigate agreement in intraocular pressure (IOP) measurements among Icare rebound tonometry, Tonopen tonometry, and Goldmann applanation tonometry following vitreoretinal surgery. DESIGN Reliability analysis of tonometers. METHODS Fifty-eyes of 50 adults undergoing vitreoretinal surgery were enrolled. IOP was measured on first postoperative day using Icare (Tiolat, Helsinki, F...

متن کامل

Comparison of intraocular pressure measurement between rebound, non-contact and Goldmann applanation tonometry in treated glaucoma patients.

BACKGROUND To compare the intraocular pressure readings obtained with the iCare rebound tonometer and the 7CR non-contact tonometer with those measured by Goldmann applanation tonometry in treated glaucoma patients. DESIGN A prospective, cross-sectional study was conducted in a private tertiary glaucoma clinic. PARTICIPANTS OR SAMPLES One hundred nine (54 males : 55 females) patients includ...

متن کامل

Evaluation of a new rebound tonometer for self-measurement of intraocular pressure Sub-title Self-measurement using Icare Home showed good agreement with Goldmann applanation

Title Evaluation of a new rebound tonometer for self-measurement of intraocular pressure Subtitle Self-measurement using Icare Home showed good agreement with Goldmann applanation tonometry. Icare Home self-measurement was well accepted and reliable results obtained by 3 in 4 subjects, suggesting potential use for home measurement of intraocular pressure. Number of references = 26 Number of sup...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Archivos de la Sociedad Espanola de Oftalmologia

دوره 89 7  شماره 

صفحات  -

تاریخ انتشار 2014