Electrical conductivity measurement of metal plates using broadband eddy-current and four-point methods

نویسندگان

  • Nicola Bowler
  • Yongqiang Huang
  • Y Huang
چکیده

Electrical conductivity of metal plates is measured by two distinct methods and the uncertainty associated with each method is evaluated. First, the impedance of an air-cored eddy-current coil is measured in the frequency range 100 Hz to 20 kHz. Corrections are made to account for the fact that the coil is not a pure inductor but exhibits finite resistance and capacitance in and between the windings. Then, the conductivity of brass and stainless steel plates is determined with 3 and 2% uncertainty (68% confidence level) by seeking the best fit (least-mean-square error) between experimental measurements of coil impedance and values calculated theoretically. The residual error in the fitting process is found to be the main indicator of uncertainty in the conductivity measurement. Second, four-point alternating current potential drop measurements are made on the same samples in the frequency range 1–100 Hz. Conductivity is determined from these measurements by means of a simple analytic formula, valid in a quasi-static regime, with an uncertainty approximately 0.5%. The main source of uncertainty in the four-point conductivity measurement is scatter in the voltage measurements. Both of these techniques give rise to smaller uncertainties in the measurement of conductivity than a MIZ-21A eddy-current instrument (2% and 40% for brass and stainless steel, respectively) and without the need for calibration specimens. In addition, the four-point approach is independent of magnetic permeability below a certain characteristic frequency and can be used to measure conductivity of ferrous metals. As an example, the conductivity of a spring steel plate is also determined.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

QUALITY CONTROL OF PRECIPITATION HARDENED ALUMINIUM ALLOY PARTS VIA EDDY-CURRENT NONDESTRUCTIVE EVALUATION

Precipitation hardening is the most common method in the strengthening of aluminium alloys. This method relies on the decrease of solid solubility with temperature reduction to produce fine precipitations which impede the movement of dislocations. The quality control of aluminium alloy specimens is an important concern of engineers. Among different methods, non-destructive techniques are the...

متن کامل

Electrical Impedance Imaging Using Eddy Current

Electric impedance imaging is a method of reconstructing spatial distribution of electrical conductivity inside a subject. In this paper, a new method of electrical impedance imaging using eddy current is proposed. The eddy current distribution in the body depends on the conductivity distribution and the magnetic field pattern. By changing the position of magnetic core, a set of voltage differe...

متن کامل

Evaluation of Acoustic Emission Inspection of Oil Tank Floor via Tank Bottom Plates Thickness Measurement

This paper deals with an inspection of defects including corrosion and leakage by Acoustic Emission (AE) in an oil tank floor. In contrast to intrusive methods such as magnetic flux leakage, eddy current or ultrasonic, there is no need to emptying and entering the tank in AE method in order to inspection. In this research AE investigation and AE source location is presented in detail. The resul...

متن کامل

Design of Nondestructive System for Hardness of Carbonitrided Steel Parts ‎Using Eddy Current Testing Method: Case study on the Tappet Parts

In this paper, a non-destructive eddy current test system was designed to evaluate the surface hardness of the carbonitrided steel parts. For this purpose, various samples of the tappet were used after the surface hardening and grinding operations as control samples and, if necessary, were subjected to destructive and non-destructive tests. Microstructural studies were performed with a microsco...

متن کامل

An Insight into Colloidal Gas Aphron Drainage Using Electrical Conductivity Measurement

In the present  paper Electrical Conductivity (EC) of Colloidal Gas Aphron (CGA) suspensions was measured for anionic and cationic surfactants (Sodium Dodecyl Sulfate (SDS) and Tetradecyl Trimethyl Ammonium Bromide (TTAB)). Experiments were made for different concentrations of SDS (6, 8.1, 10 mM) and TTAB (2, 3.51, 5 mM). CGA drainage behavior was observed and measured using 1 liter measuri...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره   شماره 

صفحات  -

تاریخ انتشار 2005