Prospective evaluation of patency and early experience utilizing an automated distal anastomosis device (C-port).
نویسندگان
چکیده
BACKGROUND : In November 2005, a new automated distal anastomotic device (C-Port) for coronary artery bypass grafting (CABG) was cleared by the FDA for use in the United States. This study represents a prospective evaluation of our early experience using the device with graft patency assessment determined by 64 slice multidetector row computed tomography angiography (64CT). MATERIALS AND METHODS : Patients undergoing coronary artery bypass were evaluated preoperatively by cine angiography for possible use of the C-Port system. A final decision for its use was made clinically at operation based on target vessel and vein graft specifications. Patency of the grafts was evaluated at 1 to 7 months (average 91 days) by 64CT with IRB approval and signed informed consent. RESULTS : The C-Port device was used to form 69 distal anastomoses in 50 patients (46 CABG were performed off pump). There were eight misfires with no adverse effects for an immediate success rate of 88.4%. Thirty-five of 41 patients were available postoperatively for evaluation by 64CT (85.4% follow-up). There was one postoperative death. Seventeen sequential anastomoses were excluded because of inability to reliably evaluate patency by 64CT. Four 64CT scans were uninterpretable. For the remaining 31 patients, 64CT scans were used to evaluate for graft patency. The overall patency rate for all anastomoses was 94.5% (86/91) with selected patency for internal mammary artery of 100% (28/28), for C-Port connectors of 93.3% (42/45), and for hand-sewn venous end-to-side anastomosis of 88.9% (16/18). The difference in patency between C-Port anastomoses and hand sewn was not statistically significant (P = 0.62). CONCLUSIONS : The C-Port system provides reliable and reproducible compliant automated distal venous anastomoses, which results in acceptable patency at intermediate follow-up. With proper training, the learning curve for using this connector system is relatively short and the need to abandon the procedure is low. 64CT is an excellent noninvasive method for cardiac surgeons and their institutions to accurately evaluate their actual surgical results.
منابع مشابه
Early patency evaluation of new distal anastomotic device in internal mammary artery grafts using computed tomography angiography.
INTRODUCTION : Traditional coronary artery bypass grafting is performed using a hand sewn technique. The C-Port xA and Flex A anastomotic stapling devices (Cardica Inc., Redwood City, CA) were cleared by the Food and Drug Administration for use in distal coronary anastomoses in November 2006 and April 2007, respectively. They provide the ability to create a compliant, consistently reproducible,...
متن کاملDistal anastomotic patency of the Cardica C-PORT(R) xA system versus the hand-sewn technique: a prospective randomized controlled study in patients undergoing coronary artery bypass grafting.
OBJECTIVES The C-Port® Distal Anastomosis Systems (Cardica, Inc., Redwood City, CA, USA) demonstrated favourable results in feasibility trials. However, distal vein anastomoses created with the C-Port® or C-Port xA® system have never been compared with hand-sewn distal vein anastomoses. The objective of this study was to compare distal end-to-side anastomoses facilitated with the C-Port xA® Sys...
متن کاملFive-year clinical outcome and patency rate of device-dependent venous grafts after clampless OPCAB with PAS-port automated proximal anastomosis: the PAPA Study.
OBJECTIVE To evaluate long-term clinical performance and angiographic patency of automated proximal venous anastomoses following clampless coronary artery bypass (C-CAB). METHODS Observational study in patients submitted for isolated C-CAB and at least one proximal aortosaphenous anastomosis performed with an automated connector (Cardica PAS-Port) including 152 consecutive patients (165 devic...
متن کاملEvaluation of the PAS-Port Proximal Anastomosis System in coronary artery bypass surgery (the EPIC trial).
OBJECTIVE During coronary surgery, proximal vein graft anastomoses have been performed by using an aortic partial occlusion clamp to allow for a hand-sewn anastomosis. The purpose of this multicenter, prospective, randomized trial was to evaluate the efficacy of the PAS-Port device (Cardica, Inc, Redwood City, Calif), which allows an automated proximal anastomosis to be performed without aortic...
متن کاملTwelve-month patency with the PAS-port proximal connector device: a single center prospective randomized trial. Invited commentary.
BACKGROUND The PAS-Port (Cardica Inc, Redwood City, CA) is an automated system that allows for the clampless anastomosis of vein grafts to the aorta. The intent of this study was to prospectively compare one year graft patency of this system with conventional hand-sewn anastomoses in a prospective, randomized trial. METHODS A total of 99 patients undergoing elective off-pump coronary bypass s...