The effect of body contouring on the dose distribution delivered with volumetric‐modulated arc therapy technique
نویسندگان
چکیده
The purpose of the study was to investigate the dosimetric effect defining the body structure with various Hounsfield unit (HU) threshold values on the dose distributions of volumetric-modulated arc therapy (VMAT) plans. Twenty patients with prostate cancer and twenty patients with head and neck (H&N) cancer were retrospectively selected. For each patient, the body structure was redefined with HU threshold values of -180 (Body180), -350 (Body350), -700 (Body700), and -980 (Body980). For each patient, dose-volumetric parameters with those body structures were calculated using identical VMAT plans. The differences in dose-volumetric parameters due to the varied HU threshold values were calculated. For the prostate boost target volume, the maximum dose, mean dose, D95%, and D5% with Body180 were higher than those with Body980 by approximately 0.7% (p < 0.001). For H&N target volumes, the changes in D95% of the targets receiving 67.5 Gy, 54 Gy, and 48 Gy between Body180 and Body980 were -1.2%, -0.9%, and -1.2%, respectively (p < 0.001). The differences were larger for H&N VMAT plans than for prostate VMAT plans due to the inclusion of an immobilization device in the irradiated region in H&N cases. To apply all attenuating materials to dose calculation, the body structure would be defined with -980 HU. Otherwise, systematic error of about 1%, resulting in underdosage of the target volume, can occur.
منابع مشابه
Quantification and modelling of the dosimetric impact of the treatment couch in volumetric modulated arc therapy (VMAT)
Background: As the volumetric modulated arc therapy (VMAT) becoming a main role of treatment ways, the effect of couch top becomes more significant. It is imperative to re-evaluate the couches that previously may have been considered of no importance during early treatment techniques. The impact of couch top on radiation delivery was explored and the couch model was tested with the aim of reduc...
متن کاملEarly toxicity of moderate hypofractionated volumetric modulated Arc radiotherapy for localized prostate cancer
Background: Based on the radiation biology model of prostate cancer, hypofractionated radiotherapy can improve the treatment outcomes without increasing toxicity. Although hypofractionated radiotherapy is implemented over a short period of time, it is more convenient and cheaper compared with conventional fractionated treatment. The aim of this study was to investigate the early toxicity of mod...
متن کاملDosimetric Evaluation of Volumetric Modulated Arc Therapy (VMAT) and Intensity Modulated Radiotherapy (IMRT) Using AAPM TG 119 Protocol
Background: The commissioning accuracy of Volumetric Modulated Arc Therapy (VMAT) need to be evaluated.Objective: To test and evaluate commissioning accuracy of VMAT based on the TG 119 protocols at local institution. Material and Methods: The phantom, structure sets, VMAT and IMRT beam parameter setup, dose prescriptions and planning objectives were following TG 119 guidelines to c...
متن کاملOptimization of prostate cancer radiotherapy using of a spacer gel, volumetric modulated arc therapy and a single biological organ at risk objective
Background: The aim was to evaluate the benefit of technical advances for treatment planning: introduction of a hydrogel spacer, VMAT (volumetric modulated arc therapy) and a single biological organ at risk objective for the rectum and bladder. Initial standard was a step-and-shoot IMRT (intensity modulated radiotherapy) without a spacer and conventional organ at risk objectives. Material...
متن کاملDosimetric influence of Flattening Filter (FF) and Flattening Filter Free (FFF) 6 and 10 MV photon beams on Volumetric Modulated Arc Therapy (VMAT) planning in case of prostate carcinoma
Background: In the treatment of prostate cancer, radiotherapy is the potential to increase second primary cancers such as bladder and rectal cancers. The reasons for this potential are more monitor units (MUs), therefore a larger total body dose because of leakage radiation, a bigger volume of normal tissue is exposed to lower radiation doses. This study was designed to compare the integral dos...
متن کامل