Capsular Contraction with S-Shaped Deformity of Nonlength-Expanding Inflatable Penile Prosthesis Cylinders: Management and Prevention Strategies

نویسندگان

  • Edward Karpman
  • Gerard Henry
چکیده

INTRODUCTION Capsular contraction (CC) occurring with inflatable penile prosthesis (IPP) reservoirs has been reported by urologists as a cause of autoinflation. The concept of CC occurring around IPP cylinders has not been studied. AIMS Herein we report a case of CC occurring with nonlength-expanding IPP cylinders, resulting in an S-shaped deformity. MAIN OUTCOME MEASURES We sought to report a novel cause of S-shaped deformity in patients after IPP surgery. METHODS We reviewed a recent clinical case and conducted a literature review on capsular scar formation in prosthetic surgery. We also conducted a literature review of the inflammatory cascade related with prosthetic surgery. RESULTS Capsular contracture of nonlength-expanding cylinders resulting in S-shaped deformity has not been previously reported. The role of certain inflammatory markers seems to play a common role of capsular contracture in the penis and other prosthetic implantation sites. CONCLUSIONS Capsular contractures around the cylinders of an IPP can cause deformity, even in patients who have appropriately sized, nonlength-expanding cylinders. A better understanding is needed regarding the mechanical properties of the tunica albuginea and the inflammatory cascade associated with penile implant surgery. This case represents an argument for early and aggressive postoperative inflation of the IPP. Future studies should evaluate the role of inflammation modulators as adjuvant therapy after IPP surgery. Karpman E and Henry G. Capsular contraction with S-shaped deformity of nonlength-expanding inflatable penile prosthesis cylinders: Management and prevention strategies. Sex Med 2013;1:95-98.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Vacuum Preparation, Optimization of Cylinder Length and Postoperative Daily Inflation Reduces Complaints of Shortened Penile Length Following Implantation of Inflatable Penile Prosthesis

Introduction: The inflatable penile prosthesis (IPP) has been used to treat erectile dysfunction for 40 years. Loss of penile length following IPP remains the single biggest patient complaint. We describe a preoperative and postoperative patient preparation protocol to assist in setting realistic patient expectations and decreasing the complaint of reduced penile length. Materials & Methods: 75...

متن کامل

Erosion of an inflatable penile prosthesis due to autoinflation.

A 59-year-old man presented with erosion of an inflatable penile prosthesis. The cylinders were not deflated and the right one protruded from the urethra. Autoinflation that causes failure of deflation is an annoying and embarrassing mechanical failure. This failure may cause intraurethral erosion of the prosthesis. Urologists should inform their patients who hope for implantation of an inflata...

متن کامل

Mechanism of Penile Prosthesis Induced Urethral Stricture: Treatment and Prevention

The advent of phosphodiesterase type 5 inhibitor has revolutionized the management of erectile dysfunction (ED). Since the introduction of sildenafil citrate in 1998, oral systemic therapy has become the first line therapy for men with ED. However, oral pharmacotherapy often fails in patients with severe diabetes, radical prostatectomy, Peyronie’s disease, and severe penile fibrosis. These pati...

متن کامل

AB022. New strategy for inflatable penile prosthesis reservoir placement and removal

Inflatable penile prostheses (IPP) have been utilized reliably as a definitive treatment for medically refractory erectile dysfunction (ED) over the last four decades. The device consists of two intra-corporal penile cylinders, a pump placed in the scrotum, and a fluidcontaining reservoir. However, placement of an IPP reservoir in a patient with multiple lower abdominal/ pelvic surgeries is tec...

متن کامل

Fracture at the input tube-cylinder junction of AMS 700 inflatable penile prostheses as a complication of a modified implantation technique in a series of 99 patients.

OBJECTIVES To compare the incidence of a specific failure mode of the penile implant, fracture at the input tube-cylinder junction, with respect to two methods of managing the input tube. METHODS AMS 700 series three-piece inflatable penile prostheses were implanted in the first 26 patients using an ordinary technique in which the input tubing runs alongside the cylinder within the corpus and...

متن کامل