Clinical performance evaluation of total protein measurement by digital refractometry and characterization of non-protein solute interferences
نویسندگان
چکیده
OBJECTIVES Refractometric methods to measure total protein (TP) in serum and plasma specimens have been replaced by automated biuret methods in virtually all routine clinical testing. A subset of laboratories, however, still report using refractometry to measure TP in conjunction with serum protein electrophoresis. The objective of this study was therefore to conduct a modern performance evaluation of a digital refractometer for TP measurement. DESIGN AND METHODS Performance evaluation of a MISCO Palm Abbe™ digital refractometer was conducted through device familiarization, carryover, precision, accuracy, linearity, analytical sensitivity, analytical specificity, and reference interval verification. Comparison assays included a manual refractometer and an automated biuret assay. RESULTS Carryover risk was eliminated using a demineralized distilled water (ddH2O) wash step. Precision studies demonstrated overall imprecision of 2.2% CV (low TP pool) and 0.5% CV (high TP pool). Accuracy studies demonstrated correlation to both manual refractometry and the biuret method. An overall positive bias (+5.0%) was observed versus the biuret method. On average, outlier specimens had an increased triglyceride concentration. Linearity was verified using mixed dilutions of: a) low and high concentration patient pools, or b) albumin-spiked ddH2O and high concentration patient pool. Decreased recovery was observed using ddH2O dilutions at low TP concentrations. Significant interference was detected at high concentrations of glucose (>267 mg/dL) and triglycerides (>580 mg/dL). Current laboratory reference intervals for TP were verified. CONCLUSIONS Performance characteristics of this digital refractometer were validated in a clinical laboratory setting. Biuret method remains the preferred assay for TP measurement in routine clinical analyses.
منابع مشابه
Evaluation and analytical validation of a handheld digital refractometer for urine specific gravity measurement
OBJECTIVES Refractometers are commonly used to determine urine specific gravity (SG) in the assessment of hydration status and urine specimen validity testing. Few comprehensive performance evaluations are available demonstrating refractometer capability from a clinical laboratory perspective. The objective of this study was therefore to conduct an analytical validation of a handheld digital re...
متن کاملEvaluation of Single Cell Protein as a Non-Conventional Feedstuff in Broilers Feeding
Single cell protein (SCP) is the definition for dried cells originated from single-celled organisms intended to be used as a protein source in human foods or animal feeds. An experiment has been conducted to evaluate the various levels of single cell protein on fecal microflora and productive traits of broiler chickens. A total of 192 day-old Ross 308 male chicks were randomly divided into 4 tr...
متن کاملCapillarys 3 Octa®: Analytical Performance Assessment for HbA1c Quantification
Objective: Determination of HbA1c level is a precious indicator for therapeutic follow-up of patients with type 1 or type 2 diabetes. Our study aimed to evaluate the analytical characteristics of Capillarys 3 Octa® HbA1c measurement by capillary electrophoresis. Materials and Methods: Our study involved 265 venous whole blood specimens repeatability, intermediate fidelity, accuracy, linearity,...
متن کاملComments on "experimental determination of net protein charge, [A]tot, and Ka of nonvolatile buffers in bird plasma".
TO THE EDITOR: I am writing to express some concerns about the article by Stämpfli et al. (7), “Experimental determination of net protein charge, [A]tot, and Ka of nonvolatile buffers in bird plasma.” First, it appears that the BCG dye binding technique was used to determine albumin concentration. This technique has not been validated for avian albumin and does not correlate with electrophoreti...
متن کاملDiagnostic Value of Measurement Specific Gravity by Refractometric and Dipstick Method in Differentiation between Transudate and Exudate in Pleural and Peritoneal Fluid
Background: Accumulation of pleural and peritoneal fluid is seen in some diseases. In order to diagnose the disease and start the treatment, one of the most important actions will be to differentiate between exudates and transudates. The objective of this study was to determine the diagnostic value of measuring the specific gravity of the fluid through...
متن کامل