FREE ABSTRACTS Comparing Cochlear Implant Users' Speech Performance with Processor Fittings Based on Conventionally Determined T and C Levels or on Compound Action Potential Thresholds and Live- Voice Speech in a Prospective Balanced Crossover

نویسندگان

  • Christina Willeboer
  • Guido F. Smoorenburg
  • Navid Shahnaz
  • Karin Bork
چکیده

Objective: The objective of the present study is to improve the efficiency of the fitting procedure of cochlear implant processors by making use of measurements of the electrically evoked compound actio potential (ECAP) and live-voice speech. Design: In a randomised prospective cross-over design we compare speech performance of eighteen adult subjects when following the conventional fitting procedure to a procedure in which we use the profile of the ECAP threshold levels across the full electrode array measured intra-operatively. The overall level of the profile is shifted (by an equal amount of current units per electrode) until we find the threshold for live speech (new T levels) and the loudness comfort level (new C levels). Each fitting procedure is tested for 6 wk. Speech performance is measured in quiet and in noise every other week. Results: The results show little difference between the scores (Dutch CVC words) for the conventional fitting procedure and the ECAP based fitting, although the T and C levels may differ markedly. Conclusion: The new fitting procedure is much faster and easier in the initial phase. Further improvement of performance may be obtained in a later stage of the fitting procedure by changing some individual electrodes on the basis of subjective responses. Wideband Reflectance Norms for Caucasian and Chinese Young Adults. Ear & Hearing. 27(6):774-788, December 2006. Navid Shahnaz; Karin Bork Abstract: Objective: This study examined differences between the middle ears of two ethnic groups, Caucasian and Chinese young adults with normal hearing, using a new middle-ear measurement technique, wideband energy reflectance. The goal of this study was to determine whether the Chinese group had different middleear transmission properties than the Caucasian group. Objective: This study examined differences between the middle ears of two ethnic groups, Caucasian and Chinese young adults with normal hearing, using a new middle-ear measurement technique, wideband energy reflectance. The goal of this study was to determine whether the Chinese group had different middleear transmission properties than the Caucasian group. Design: There were 126 subjects (237 ears) between the ages of 18 and 32 yr, with 62 subjects in the Caucasian group and 64 subjects in the Chinese group. Wideband energy reflectance data were gathered using Mimosa Acoustics (RMS system version 4.0.4.4) wideband reflectance (WBR) equipment. Results: The Chinese group had significantly lower wideband energy reflectance than their Caucasian counterparts at higher frequencies; however, the Caucasian group had significantly lower energy reflectance at lower frequencies than the Chinese group. The Chinese group also had significantly lower admittance magnitude than the Caucasian group at lower frequencies. Because body size indices are more comparable between Caucasian females and Chinese males, the effect of body size could be potentially adjusted for by comparing the Caucasian female subjects with the Chinese male subjects. The differences observed between the Caucasian and the Chinese groups were no longer significant when the Caucasian female subjects were compared with the Chinese male subjects. Applying the Caucasian norms to four Caucasian adults with surgically confirmed otosclerosis resulted in an improved hit rate compared with the combined Caucasian and Chinese norms and the Chinese-only norms.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Relationship between Electrically Evoked Compound Action Potential Thresholds and Auditory, Language, and Speech Progress after Cochlear Implant Surgery

Introduction: Electrically evoked compound action potential (ECAP) is an objective auditory response that can be used in the programing of cochlear implants. The aims of this study were to monitor ECAP thresholds and auditory, language and speech progress for 6 months after cochlear implant surgery and to evaluate any relationship between them.   Materials and Methods: Ten children with a m...

متن کامل

Evaluation of the Slope of Amplitude Growth Function Changes of the Electrically Evoked Action Potential in Three Months after Receiving the Device in Children with Cochlear Implant

Objective: In neural response telemetry, intracochlear electrodes stimulate the auditory nerve and record the neural responses. The electrical stimulation is sent to the auditory nerve by an electrode and the resulted response, called electrically evoked compound action potential, is recorded by an adjacent electrode. The most important clinical applications of this test are evaluation and moni...

متن کامل

Comparison of electrically evoked compound action potential thresholds and loudness estimates for the stimuli used to program the Advanced Bionics cochlear implant.

BACKGROUND In the mid-1990s, Cochlear Corporation introduced a cochlear implant (CI) to the market that was equipped with hardware that made it possible to record electrically evoked compound action potentials (ECAPs) from CI users of all ages. Over the course of the next decade, many studies were published that compared ECAP thresholds with levels used to program the speech processor of the Nu...

متن کامل

Influence of Evoked Compound Action Potential on Speech Perception in Cochlear Implant Users Summary

El ectrically Evoked Compound Action Potential is a measure of synchronous cochlear nerve fibers activity elicited by electrical stimulation of the cochlear implant. The electrophysiological nerve responses may contribute to explain the variability in individual performance of cochlear implant recipients. Aim: To compare speech perception tests' performances of cochlear implant users according ...

متن کامل

Influence of evoked compound action potential on speech perception in cochlear implant users.

UNLABELLED Electrically Evoked Compound Action Potential is a measure of synchronous cochlear nerve fibers activity elicited by electrical stimulation of the cochlear implant. The electrophysiological nerve responses may contribute to explain the variability in individual performance of cochlear implant recipients. AIM To compare speech perception tests performances of cochlear implant users ...