A dosimetric comparison of volumetric modulated arc therapy (VMAT) and non-coplanar intensity modulated radiotherapy (IMRT) for nasal cavity and paranasal sinus cancer

نویسندگان

  • Yuri Jeong
  • Sang-wook Lee
  • Jungwon Kwak
  • Ilsung Cho
  • Sang Min Yoon
  • Jong Hoon Kim
  • Jin-hong Park
  • Eun Kyung Choi
  • Si Yeol Song
  • Young Seok Kim
  • Su Ssan Kim
  • Ji Hyeon Joo
  • Seung Do Ahn
چکیده

BACKGROUND To compare dosimetric parameters of volumetric modulated arc therapy (VMAT) and non-coplanar intensity modulated radiotherapy (IMRT) for nasal cavity and paranasal sinus cancer with regard to the coverage of planning target volume (PTV) and the sparing of organs at risk (OAR). METHODS Ten patients with nasal cavity or paranasal sinus cancer were re-planned by VMAT (two-arc) plan and non-coplanar IMRT (7-, 11-, and 15-beam) plans. Planning objectives were to deliver 60 Gy in 30 fractions to 95% of PTV, with maximum doses (D(max)) of <50 Gy to the optic nerves, optic chiasm, and brainstem, <40 Gy to the eyes and <10 Gy to the lenses. The target mean dose (D(mean)) to the parotid glands was <25 Gy, and no constraints were applied to the lacrimal glands. Planning was optimized to minimized doses to OAR without compromising coverage of the PTV. VMAT and three non-coplanar IMRT (7-, 11-, and 15-beam) plans were compared using the heterogeneity and conformity indices (HI and CI) of the PTV, D(max) and D(mean) of the OAR, treatment delivery time, and monitor units (MUs). RESULTS The HI and CI of VMAT plan were superior to those of the 7-, 11-, and 15-beam non-coplanar IMRT. VMAT and non-coplanar IMRT (7-, 11-, and 15-beam) showed equivalent sparing effects for the optic nerves, optic chiasm, brainstem, and parotid glands. For the eyes and lenses, VMAT achieved equivalent or better sparing effects when compared with the non-coplanar IMRT plans. VMAT showed lower MUs and reduced treatment delivery time when compared with non-coplanar IMRT. CONCLUSIONS In 10 patients with nasal cavity or paranasal sinus cancer, a VMAT plan provided better homogeneity and conformity for PTV than non-coplanar IMRT plans, with a shorter treatment delivery time, while achieving equal or better OAR-sparing effects and using fewer MUs.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Dosimetric Evaluation of Volumetric Modulated Arc Therapy (VMAT) and Intensity Modulated Radiotherapy (IMRT) Using AAPM TG 119 Protocol

Background: The commissioning accuracy of Volumetric Modulated Arc Therapy (VMAT) need to be evaluated.Objective: To test and evaluate commissioning accuracy of VMAT based on the TG 119 protocols at local institution. Material and Methods: The phantom, structure sets, VMAT and IMRT beam parameter setup, dose prescriptions and planning objectives were following TG 119 guidelines to c...

متن کامل

Optimization of prostate cancer radiotherapy using of a spacer gel, volumetric modulated arc therapy and a single biological organ at risk objective

Background: The aim was to evaluate the benefit of technical advances for treatment planning: introduction of a hydrogel spacer, VMAT (volumetric modulated arc therapy) and a single biological organ at risk objective for the rectum and bladder. Initial standard was a step-and-shoot IMRT (intensity modulated radiotherapy) without a spacer and conventional organ at risk objectives.&nbsp; Material...

متن کامل

Dosimetric comparison of intensity-modulated radiotherapy (IMRT) and volumetric modulated arc therapy (VMAT) in total scalp irradiation: a single institutional experience

PURPOSE Total scalp irradiation (TSI) is a rare but challenging indication. We previously reported that non-coplanar intensity-modulated radiotherapy (IMRT) was superior to coplanar IMRT in organ-at-risk (OAR) protection and target dose distribution. This consecutive treatment planning study compared IMRT with volumetric-modulated arc therapy (VMAT). MATERIALS AND METHODS A retrospective trea...

متن کامل

Non-coplanar volumetric modulated arc therapy for irradiation of paranasal sinus tumors

Background: Sinonasal cancer (SNC) arises in the proximity of many critical organs at risk (OAR). Therefore, minimizing radiation doses to critical organs is a priority. We evaluated the use of a fast, non-coplanar volumetric arc therapy (VMAT) delivery technique for this purpose.

متن کامل

Comparison of toxicity and outcome in stage III NSCLC patients treated with IMRT or VMAT

Purpose or Objective: Intensity-modulated radiation therapy (IMRT) and volumetric-modulated arc therapy (VMAT) are widely used in the treatment of advanced stage non-small cell lung cancer (NSCLC). These techniques deliver conformal dose distributions at the cost of increased target dose heterogeneity (particularly IMRT) and larger volumes of surrounding healthy tissues receiving low doses (par...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 9  شماره 

صفحات  -

تاریخ انتشار 2014