Biological effect of dose distortion by fiducial markers in spot-scanning proton therapy with a limited number of fields: a simulation study.
نویسندگان
چکیده
PURPOSE In accurate proton spot-scanning therapy, continuous target tracking by fluoroscopic x ray during irradiation is beneficial not only for respiratory moving tumors of lung and liver but also for relatively stationary tumors of prostate. Implanted gold markers have been used with great effect for positioning the target volume by a fluoroscopy, especially for the cases of liver and prostate with the targets surrounded by water-equivalent tissues. However, recent studies have revealed that gold markers can cause a significant underdose in proton therapy. This paper focuses on prostate cancer and explores the possibility that multiple-field irradiation improves the underdose effect by markers on tumor-control probability (TCP). METHODS A Monte Carlo simulation was performed to evaluate the dose distortion effect. A spherical gold marker was placed at several characteristic points in a water phantom. The markers were with two different diameters of 2 and 1.5 mm, both visible on fluoroscopy. Three beam arrangements of single-field uniform dose (SFUD) were examined: one lateral field, two opposite lateral fields, and three fields (two opposite lateral fields + anterior field). The relative biological effectiveness (RBE) was set to 1.1 and a dose of 74 Gy (RBE) was delivered to the target of a typical prostate size in 37 fractions. The ratios of TCP to that without the marker (TCP(r)) were compared with the parameters of the marker sizes, number of fields, and marker positions. To take into account the dependence of biological parameters in TCP model, α∕β values of 1.5, 3, and 10 Gy (RBE) were considered. RESULTS It was found that the marker of 1.5 mm diameter does not affect the TCPs with all α∕β values when two or more fields are used. On the other hand, if the marker diameter is 2 mm, more than two irradiation fields are required to suppress the decrease in TCP from TCP(r) by less than 3%. This is especially true when multiple (two or three) markers are used for alignment of a patient. CONCLUSIONS It is recommended that 1.5-mm markers be used to avoid the reduction of TCP as well as to spare the surrounding critical organs, as long as the markers are visible on x-ray fluoroscopy. When 2-mm markers are implanted, more than two fields should be used and the markers should not be placed close to the distal edge of any of the beams.
منابع مشابه
Biological effect of dose distortion by fiducial markers in spot-scanning proton therapy with a limited number of fields: a simulation study Running head: Biological effect of proton dose-distortion by fiducials
Title Biological effect of dose distortion by fiducial markers in spot-scanning proton therapy with a limited number of fields: A simulation study Author(s) Matsuura, Taeko; Maeda, Kenichiro; Sutherland, Kenneth; Takayanagi, Taisuke; Shimizu, Shinichi; Takao, Seishin; Miyamoto, Naoki; Nihongi, Hideaki; Toramatsu, Chie; Nagamine, Yoshihiko; Fujimoto, Rintaro; Suzuki, Ryusuke; Ishikawa, Masayori;...
متن کاملA Proton Beam Therapy System Dedicated to Spot-Scanning Increases Accuracy with Moving Tumors by Real-Time Imaging and Gating and Reduces Equipment Size
PURPOSE A proton beam therapy (PBT) system has been designed which dedicates to spot-scanning and has a gating function employing the fluoroscopy-based real-time-imaging of internal fiducial markers near tumors. The dose distribution and treatment time of the newly designed real-time-image gated, spot-scanning proton beam therapy (RGPT) were compared with free-breathing spot-scanning proton bea...
متن کاملInvestigation of tumor motion influence on applied dose distribution in conventional proton therapy vs. IMPT a 4D Monte Carlo simulation study
Background: in radiation treatment of moving targets located in thorax region of patient body, the delivered dose does not match with the planned treatment, resulting in some over and under dosage in the tumor volume, as a function of motion magnitude and frequency. Several efforts have been done to investigate the target motion effects on dose distribution in the target and surrounding normal ...
متن کاملEvaluation of the effective dose during PBFT for brain cancer: A Monte Carlo Study
Introduction: Recently, an approach exploiting the proton therapy biological enhancement by using Boron atoms injected inside a tumor, has been proposed. Three alpha particles with an average energy around 4MeV are emitted from the point of reaction between a proton and boron. In addition, the 719 keV prompt gamma emitted by the proton Boron fusion reactions can be used for on-...
متن کاملEffects of Defining Realistic Compositions of the Ocular Melanoma on Proton Therapy
Background: Recent studies in eye plaque brachytherapy have shown a considerable difference between the dosimetric results using water phantom and a model of human eye containing realistic materials. In spite of this fact, there is a lack of simulation studies based on such a model in proton therapy literatures. In the presented work, the effect of utilizing an eye model with ocular media on pr...