Analysis of violet-excited fluorochromes by flow cytometry using a violet laser diode.
نویسندگان
چکیده
BACKGROUND Low power violet laser diodes (VLDs) have been evaluated as potential replacements for water-cooled argon-ion and krypton-ion ultraviolet and violet lasers for DNA content analysis using the Hoechst dyes and 4,6-diamidino-2-phenylindole (Shapiro HMN, Perlmutter NG: Cytometry 44:133-136, 2001). In this study, we used a VLD to excite a variety of violet-excited fluorescent molecules important in biomedical analysis, including the fluorochromes Cascade Blue and Pacific Blue, the expressible fluorescent protein cyan fluorescent protein (CFP), and the fluorogenic alkaline phosphatase (AP) substrate 2-(5'-chloro-2'-phosphoryloxyphenyl)-6-chloro-4-(3H)-quinazoline (ELF-97; for endogenous AP detection and cell surface labeling with AP-conjugated antibodies). METHODS Comparisons were made between VLD excitation and a krypton-ion laser emitting at 407 nm (both at higher power levels and with the beam attenuated at levels approximating the VLD) on the same FACSVantage SE stream-in-air flow cytometer. We evaluated a Power Technology 408-nm VLD (30 mW) equipped with circularization optics (18 mW maximum output, set to 15 mW) and a Coherent I-302C krypton-ion laser emitting at power levels ranging from 15 to 75 mW. RESULTS Cascade Blue, Pacific Blue, and CFP showed comparable signal-to-noise ratios and levels of sensitivity with VLD excitation versus the krypton-ion laser at high and VLD-matched power outputs. Multicolor fluorescent protein analysis with 488-nm excitation of green fluorescent protein and DsRed and VLD excitation of CFP was therefore feasible and was demonstrated. Similar levels of excitation efficiency between krypton-ion and VLD sources also were observed for ELF-97 detection. CONCLUSIONS These evaluations confirmed that VLDs may be cost- and maintenance-effective replacements for water-cooled gas lasers for applications requiring violet excitation in addition to DNA binding dyes.
منابع مشابه
Analysis of UV-excited fluorochromes by flow cytometry using near-ultraviolet laser diodes.
INTRODUCTION Violet laser diodes have become common and reliable laser sources for benchtop flow cytometers. While these lasers are very useful for a variety of violet and some ultraviolet-excited fluorochromes (e.g., DAPI), they do not efficiently excite most UV-stimulated probes. In this study, the next generation of InGaN near-UV laser diodes (NUVLDs) emitting in the 370-375-nm range have be...
متن کاملViolet laser diodes in flow cytometry: an update.
INTRODUCTION In previous studies we and others have demonstrated the usefulness of violet laser diodes (VLDs) as replacement laser sources for krypton-ion lasers on stream-in-air cytometers. Previously available VLDs had a maximum available power of less than 25 mW; this was sufficient for excitation of densely labeled cell surface antigens using fluorochromes such as Cascade Blue or Pacific Bl...
متن کاملDiscrimination of the Hoechst side population in mouse bone marrow with violet and near-ultraviolet laser diodes.
BACKGROUND Discrimination of stem cells with flow cytometric analysis of Hoechst 33342 efflux by the ABCG2 transporter (termed the Hoechst side population, or SP technique) is a valuable methodology for identifying bone marrow progenitors enriched with stem cells. Unfortunately, it requires a ultraviolet (UV) laser source, usually necessitating an expensive and maintenance-intensive argon- or k...
متن کاملA simple and effective method for differentiating GFP and YFP by flow cytometry using the violet laser.
GFP and YFP are fluorescent proteins that are widely used in various cellular assays (1). The 488-nm blue laser, that is standard on virtually all flow cytometers, efficiently excites both proteins. GFP and YFP signals can be discriminated using a customized filter configuration involving a 550/30-nm band pass for YFP, a 510/20-nm band pass filter for GFP, with a 525-nm short-pass dichroic mirr...
متن کاملDPSS yellow-green 561-nm lasers for improved fluorochrome detection by flow cytometry.
INTRODUCTION Blue-green 488-nm laser sources are widespread in flow cytometry but suffer some drawbacks for cell analysis, including their excitation of endogenous proteins (resulting in high cellular autofluorescence) and their less-than-optimal coincidence with the excitation maxima of commonly used fluorochromes, including the phycoerythrins (PE). Longer wavelength lasers such as green heliu...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Cytometry. Part A : the journal of the International Society for Analytical Cytology
دوره 54 1 شماره
صفحات -
تاریخ انتشار 2003