Blood irradiation with accelerator produced electron beams.

نویسندگان

  • M J Butson
  • T Cheung
  • P K Yu
  • M J Stokes
چکیده

Blood and blood products are irradiated with gamma rays to reduce the risk of graft versus host disease (GVHD). A simple technique using electron beams produced by a medical linear accelerator has been studied to evaluate irradiation of blood and blood products. Variations in applied doses for a single field 20 MeV electron beam are measured in a phantom study. Doses have been verified with ionization chambers and commercial diode detectors. Results show that the blood product volume can be given a relatively homogeneous dose to within 6% using 20 MeV electrons without the need to rotate the blood bags or the beam entry point. The irradiation process takes approximately 6.5 minutes for 30 Gy applied dose to complete as opposed to 12 minutes for a dual field x-ray field irradiation at our centre. Electron beams can be used to satisfactorily irradiate blood and blood products in a minimal amount of time.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Monte Carlo Simulation of Electron Beams produced by LIAC Intraoperative Radiation Therapy Accelerator

Background: One of the main problems of dedicated IORT accelerators is to determine dosimetric characteristics of the electron beams. Monte Carlo simulation of IORT accelerator head and produced beam will be useful to improve the accuracy of beam dosimetry.Materials and Methods: Liac accelerator head was modeled using the BEAMnrcMonte Carlo simulation system. Phase-space files were generated at...

متن کامل

Spectrally peaked electron beams produced via surface guiding and acceleration in femtosecond laser-solid interactions.

Highly collimated MeV electron beam guiding has been observed along the target surface following the interaction of bulk target irradiation by femtosecond laser pulses at relativistic intensities. The beam quality is shown to depend critically on the laser contrast: With a ns prepulse, the generated electron beam is well concentrated and intense, while a high laser contrast produces divergent e...

متن کامل

Use of a blood glucose meter for radiochromic film analysis in blood irradiation.

The use of a diabetic blood glucose meter for radiochromic film dosimetry in blood irradiation using x-ray beams on a medical linear accelerator has been investigated. The glucose meter provides optical density analysis in the visible and infrared region using a reflectance measurement technique. By comparing the 'blood sugar' level output with standard calibration gafchromic films a calibratio...

متن کامل

Nuclear Physics Comparison of Measured and Monte Carlo Calculated Dose Distributions for the Novac7® Linear Accelerator

Percentage depth doses (PDDs) and transverse dose profiles (TDPs) were calculated by a Monte Carlo simulation of electron beams of different energies (3, 5, 7 and 9 MeV) produced by a NOVAC7® accelerator. The NOVAC7® is a linear accelerator designed for Intraoperative Radiation Therapy (IORT). The PDDs and TDPs were also experimentally determined. Both calculations and measurements were done on...

متن کامل

کاهش آلاینده‌های NOx و SOx در واحدهای صنعتی با استفاده ازشتاب دهنده الکترون

Electron beam accelerators technology has made significant progress in environmental applications in recent years. Including some environmental applications of industrial accelerators, is clearing the air from oxides of nitrogen and sulfur(NOx , SOx) produced by industrial facilities. Combustion  of coal, oil, natural gas and other gases that are produced in power plant, refineries and industri...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Physics in medicine and biology

دوره 45 11  شماره 

صفحات  -

تاریخ انتشار 2000