Metal-on-metal total hip arthroplasty: causes and high incidence of early failure.
نویسندگان
چکیده
Early failures of metal-on-metal total hip arthroplasty (THA) occur due to aseptic loosening, metal hypersensitivity reactions, pseudotumor formation, and component seizing. The purpose of this study was to investigate the timing, common modes of failure, clinical outcomes, and incidence of metal-on-metal THA revisions. A review was performed of 80 patients who underwent revision of a failed metal-on-metal THA for any reason. The most common reason for metal-on-metal failure was aseptic acetabular loosening, with a rate of 56.25% (45/80 patients). Early failure of metal-on-metal THAs was noted, with 78% of these revisions being performed within 2 years of the index operation and 92.5% within 3 years. Furthermore, 13% of patients experienced significant localized soft tissue reactions. Mean preoperative Harris Hip Score was 42.35 ± 14.24 and mean postoperative Harris Hip Score was 66.5 ± 23.2 (range, 9.55-95.4), with an average follow-up of 438 ± 492 days (range, 40-2141), or 1.2 years.It is imperative that clinicians be cognizant of the fact that the proposed advantages of metal-on-metal THA are not without potential detrimental sequelae. This article proposes an algorithm to aid in diagnosing the etiology of a painful metal-on-metal THA, as well as 2 classification schemes regarding metal-on-metal THA complications to help direct treatment.
منابع مشابه
Fatal Cobalt Cardiomyopathy Following Revision Total Hip Arthroplasty – A Brief Report with Review of Literature
Ceramic bearing surfaces are being increasingly used in young patients undergoing total hip arthroplasty. However,failures have been reported including fractures even with the newer third generation ceramics. The recommendedtreatment for fracture of ceramic bearing surfaces is complete synovectomy and revision total hip arthroplasty.However, disappointing results have also bee...
متن کاملThe clinical performance of metal-on-metal as an articulation surface in total hip replacement.
The metal-on-metal articulations in total hip arthroplasty (THA) were widely used between 1960 and 1975. The McKee-Farrar and other first-generation prostheses failed at a high rate because impingement caused early component loosening. The problem of early component loosening was corrected by improved component design and better manufacturing quality. Second-generation metal-on-metal total hip ...
متن کاملFailure of a metal-on-metal total hip arthroplasty from progressive osteolysis.
Ultra-high-molecular weight polyethylene (UHMWPE) wear, debris-induced osteolysis is a frequent cause of failure of total hip arthroplasty. Metal-on-metal total hip arthroplasty eliminates the generation of UHMWPE particulate debris. Although the volumetric wear of a metal-on-metal articulation may be lower than a metal-UHMWPE articulation, the number of particles may be higher. Osteolysis can ...
متن کاملMetal-on-Metal Total Hip Arthroplasty: Quality of Online Patient Information.
Metal-on-metal total hip arthroplasty (THA) has generated much attention in the media because of early failure of certain implant systems. This study assessed the quality, accuracy, and readability of online information on metal-on-metal THA. The search terms "metal-on-metal hip replacement" and "metal hip replacement" were entered into the 3 most popular search engines. Information quality was...
متن کاملEarly failures in total hip arthroplasty -- a changing paradigm.
Between 2001 and 2011, 1168 revision hip arthroplasties were reviewed for "early" failures within 5 years of the primary total hip arthroplasty (THA). 24.1% underwent revision within 5 years of index THA. Aseptic loosening, infection, instability, metallosis, and fracture were common modes of failure. In our previous report from 1986 to 2000, 33% were "early" revisions, with instability and ase...