Note: Gating Characteristics of an Elekta Radiotherapy Treatment Unit Measured with Three Types of Detector Short title: Gating Characteristics of an Elekta Radiotherapy Treatment Unit

نویسندگان

  • Philip M Evans
  • J Richard N Symonds-Tayler
  • Ruth Colgan
  • Nicky Letts
  • Carlos Sandin
چکیده

The characteristics of an Elekta Precise treatment machine with a gating interface were investigated. Three detectors were used: a Farmer ionisation chamber; a MatriXX ionisation chamber array and an in-house, single pulse-measurement ionisation chamber (IVC). Measurements were made of dosimetric accuracy, flatness and symmetry characteristics and duty cycle for a range of beam-on times and gating periods. Results were compared with a standard un-gated delivery as a reference. For all beam-on times, down to 0.5 s, dosimetric differences were below ±1 % and flatness and symmetry parameter variations were below ±1.5 %. For the shorter beam-on times the in-house detector deviated from the other two detectors, suggesting that this device should be used in conjunction with other detectors for absolute dosimetry purposes. However it was found to be useful for studying gated beam characteristics pulse by pulse.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Calculation of wedged dose distributions using an analytical method

Introduction: Wedge filters are used in radiotherapy to modify photon beam and improve dose uniformity in the target volume. Determination of wedge dose distribution is important to improve treatment planning accuracy. The utilization of treatment planning software to obtain dose distribution is not always available. In this work an analytical model has been developed to calcu...

متن کامل

Characteristics of gated treatment using an optical surface imaging and gating system on an Elekta linac

BACKGROUND Knowing the technical characteristics of gated radiotherapy equipment is crucial for ensuring precise and accurate treatment when using techniques such as Deep-Inspiration Breath-Hold and gating under free breathing. With one of the first installations of the novel surface imaging system Catalyst™ (C-RAD AB, Sweden) in connection with an Elekta Synergy linear accelerator (Elekta AB, ...

متن کامل

Does gated beam delivery impact delivery accuracy on an Elekta linac?

In this study, we evaluated the performance of an Elekta linac in the delivery of gated radiotherapy. Delivery accuracy was examined with an emphasis on the impact of using short gating windows (low monitor unit beam-on segments) or long beam hold times. The performance was assessed using a 20cm by 20cm open field with the radiation delivered using a range of beam-on and beam-off time periods. ...

متن کامل

Delivery efficiency of an Elekta linac under gated operation

In this study, we have characterized the efficiency of an Elekta linac in the delivery of gated radiotherapy. We have explored techniques to reduce the beam-on delay and to improve the delivery efficiency, and have investigated the impact of frequent beam interruptions on the dosimetric accuracy of gated deliveries. A newly available gating interface was installed on an Elekta Synergy. Gating s...

متن کامل

Evaluation of the Portal Imaging System Performance for an Elekta Precise Linac in Radiotherapy

Introduction: Electronic portal imaging devices (EPIDs) provide two- and three-dimensional planar and volumetric cone beam images to improve the accuracy of radiation treatment delivery. Periodic quality assurance (QA) of EPIDs is essential for dosimetric verification in radiotherapy. In this study, a QA program was implemented to evaluate the function of the EPID to be confident in applying co...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره   شماره 

صفحات  -

تاریخ انتشار 2010