Correlation between apical protrusion in the Scheimflug imaging and Corneal Hysteresis and Corneal Resistance factor by Ocular Response Analyzer, among refractive non-keratoconic Egyptian patients
نویسنده
چکیده
INTRODUCTION Apical protrusion in the central 4-mm ring in the Scheimflug imaging (Pentacam), both for the anterior and posterior floats as well as Corneal Hysteresis and Corneal Resistance Factor by Ocular Response Analyzer (ORA), generally are considered important predictors for post-Lasik ectasia. The aim of this work was to find out if there is a statistically significant correlation between these different predictors and their correlation with the central corneal thickness for refractive non-keratoconic Egyptian patients trying to achieve a better decision and avoiding ectasia. METHODS This case-control study involved 142 eyes (of 77 patients with various refractive errors) arriving at the refractive surgery unit in the Research Institute of Ophthalmology in Giza (Egypt) in 2014-2015 seeking excimer laser ablation. The flattest, steepest keratometry readings, central corneal thickness as well as the apical protrusion in the central 4-mm ring, both for the anterior and posterior floats, in microns were measured by Scheimflug imaging. The Corneal Hysteresis and Corneal Resistance Factor were measured by the ocular response analyzer. Statistical analysis was performed by SPSS, using the Pearson correlation test. RESULTS The spherical refractive error ranged from +7.00 to -13.00 diopters (-3.80 ± 2.89). The central pachymetry ranged from 494 to 634 μm (550.35 ± 32.13). For the central 4-mm ring, the apical protrusion ranged from 0 to +15 μ (6.93 ± 2.99) for the anterior float and from -3 to +20 μ (9.33 ± 4.55) for the posterior float. The Corneal Hysterisis (CH) ranged from 7 to 14.8 mmHg (10.18±1.44), while the Corneal Resistance Factor (CRF) ranged from 7.5 to 14.9 mmHg (10.58 ± 1.67). There was a strong positive correlation between the central corneal thickness and both Corneal Hysteresis (CH: r = 0.56, P ≤ 0.01) and Corneal Resistance Factor (r = 0.46, P ≤ 0.01). A significant correlation (P < 0.05, r = 0.15) existed between apical protrusion in the posterior float and the central corneal thickness. Also, significant negative correlation (P < 0.05, r = -0.12) existed between apical protrusion in the anterior float by pentacam and the Corneal Resistance Factor by ocular response analyzer. CONCLUSIONS Our finding of a strong positive correlation between both Corneal Hysteresis and Corneal Resistance Factor and the Central corneal thickness being important for biomechanical corneal stability. The findings of this study also support using both machines preoperatively to decrease the risk of post-Lasik ectasia.
منابع مشابه
ویژگیهای بیومکانیکی قرنیه در بیماران مبتلا به کراتوکونوس و ارتباط آن با شدت بیماری
Background and Objective: Keratoconus is a disorder in cornea characterized by constant thinning and prominence that can result in vision defect. Due to ultrasonic changes in the corneal structure caused by keratoconus disease, its rigidity and biomechanical properties are subject to some changes. In this regard, the present study was conducted to determine the correlation of biomechanical feat...
متن کاملORA waveform-derived biomechanical parameters to distinguish normal from keratoconic eyes.
PURPOSE To evaluate the ability of the Ocular Response Analyzer (ORA; Reichert Ophthalmic Instruments, Buffalo, NY) to distinguish between normal and keratoconic eyes, by comparing pressure and waveform signal-derived parameters. METHODS This retrospective comparative case series study included 112 patients with normal corneas and 41 patients with bilateral keratoconic eyes. One eye from each...
متن کاملThe Relationship between Corneal Biomechanics and Corneal Shape in Normal Myopic Eyes
Introduction In a broad review, evidence that the corneal biomechanical properties influence the results and outcomes of various ocular measurements and procedures has existed for some time [1,2]. However, assessing the biomechanical properties of corneal tissue in vivo has previously not been possible. With the recent introduction of the Reichert Ocular Response Analyzer (ORA), direct clinical...
متن کاملIntraocular Pressure Measurements by Three Different Tonometers in Children with Aphakic Glaucoma and a Thick Cornea
Background: To evaluate the agreement in intraocular pressure (IOP) measurements by Ocular Response Analyzer (ORA) and Tono-Pen XL (TXL) with the Goldmann Applanation Tonometer (GAT) and to examine corneal biomechanical properties in aphakic glaucoma patients with a central corneal thickness (CCT) >600 µ. Methods: Thirty-six eyes of aphakic glaucoma patients (group 1) and 40 eyes of normal chil...
متن کاملMultivariate Analysis of the Ocular Response Analyzer's Corneal Deformation Response Curve for Early Keratoconus Detection
Purpose. To thoroughly analyze corneal deformation responses curves obtained by Ocular Response Analyzer (ORA) testing in order to improve subclinical keratoconus detection. Methods. Observational case series of 87 control and 73 subclinical keratoconus eyes. Examination included corneal topography, tomography, and biomechanical testing with ORA. Factor analysis, logistic regression, and receiv...
متن کامل