Quantitative blood flow measurements in gliomas using arterial spin-labeling at 3T: intermodality agreement and inter- and intraobserver reproducibility study.

نویسندگان

  • T Hirai
  • M Kitajima
  • H Nakamura
  • T Okuda
  • A Sasao
  • Y Shigematsu
  • D Utsunomiya
  • S Oda
  • H Uetani
  • M Morioka
  • Y Yamashita
چکیده

BACKGROUND AND PURPOSE QUASAR is a particular application of the ASL method and facilitates the user-independent quantification of brain perfusion. The purpose of this study was to assess the intermodality agreement of TBF measurements obtained with ASL and DSC MR imaging and the inter- and intraobserver reproducibility of glioma TBF measurements acquired by ASL at 3T. MATERIALS AND METHODS Two observers independently measured TBF in 24 patients with histologically proved glioma. ASL MR imaging with QUASAR and DSC MR imaging were performed on 3T scanners. The observers placed 5 regions of interest in the solid tumor on rCBF maps derived from ASL and DSC MR images and 1 region of interest in the contralateral brain and recorded the measured values. Maximum and average sTBF values were calculated. Intermodality and intra- and interobsever agreement were determined by using 95% Bland-Altman limits of agreement and ICCs. RESULTS The intermodality agreement for maximum sTBF was good to excellent on DSC and ASL images; ICCs ranged from 0.718 to 0.884. The 95% limits of agreement ranged from 59.2% to 65.4% of the mean. ICCs for intra- and interobserver agreement for maximum sTBF ranged from 0.843 to 0.850 and from 0.626 to 0.665, respectively. The reproducibility of maximum sTBF measurements obtained by methods was similar. CONCLUSIONS In the evaluation of sTBF in gliomas, ASL with QUASAR at 3T yielded measurements and reproducibility similar to those of DSC perfusion MR imaging.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Inter-Vendor Reproducibility of Pseudo-Continuous Arterial Spin Labeling at 3 Tesla

PURPOSE Prior to the implementation of arterial spin labeling (ASL) in clinical multi-center studies, it is important to establish its status quo inter-vendor reproducibility. This study evaluates and compares the intra- and inter-vendor reproducibility of pseudo-continuous ASL (pCASL) as clinically implemented by GE and Philips. MATERIAL AND METHODS 22 healthy volunteers were scanned twice o...

متن کامل

Regional reproducibility of pulsed arterial spin labeling perfusion imaging at 3T

Arterial spin labeling (ASL) is a promising non-invasive magnetic resonance imaging (MRI) technique for measuring regional cerebral blood flow (rCBF) or perfusion in vivo. To evaluate the feasibility of ASL as a biomarker for clinical trials, it is important to examine test-retest reproducibility. We investigated both inter- and intra-session reproducibility of perfusion MRI using a pulsed ASL ...

متن کامل

Acquisition time and reproducibility of continuous arterial spin-labeling perfusion imaging at 3T.

Arterial spin-labeling (ASL) is a relatively new and noninvasive MR imaging technique, used to measure cerebral blood flow (CBF). Scanning time and reproducibility remain important issues in the clinical applicability of ASL. We expected both to benefit from higher field strengths. We describe that when performing ASL at 3T, 20 averages suffice to obtain steady and reproducible CBF values. Scan...

متن کامل

Prognostic Value of Blood Flow Measurements Using Arterial Spin Labeling in Gliomas

The period of event-free survival (EFS) within the same histopathological glioma grades may have high variability, mainly without a known cause. The purpose of this study was to reveal the prognostic value of quantified tumor blood flow (TBF) values obtained by arterial spin labeling (ASL) for EFS in patients with histopathologically proven astrocytomas independent of WHO (World Health Organiza...

متن کامل

Multi-vendor reliability of arterial spin labeling perfusion MRI using a near-identical sequence: Implications for multi-center studies

INTRODUCTION A main obstacle that impedes standardized clinical and research applications of arterial spin labeling (ASL), is the substantial differences between the commercial implementations of ASL from major MRI vendors. In this study, we compare a single identical 2D gradient-echo EPI pseudo-continuous ASL (PCASL) sequence implemented on 3T scanners from three vendors (General Electric Heal...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • AJNR. American journal of neuroradiology

دوره 32 11  شماره 

صفحات  -

تاریخ انتشار 2011