Theory and calibration of the Schi0tz tonometer II. The resistance of the cornea to deformation
نویسندگان
چکیده
.lthough the cornea is very flexible, force must be required to bend it. Goldmann and Schmidt' and Gloster and Perkins following Imbert and Fick have considered bending resistance in applantation tonometry. Friedenwald however, in his studies on Schi0tz tonometry, came to the conclusion that the resistance of the cornea to deformation is negligibly small. The corneal bending forces involved in the case of the Schi0tz tonometer are by no means simple. It is a common clinical observation that during operation for cataract the corneal dome may either retain
منابع مشابه
Theory and calibration of the Schiotz tonometer. II. The resistance of the cornea to deformation.
.lthough the cornea is very flexible, force must be required to bend it. Goldmann and Schmidt' and Gloster and Perkins following Imbert and Fick have considered bending resistance in applantation tonometry. Friedenwald however, in his studies on Schi0tz tonometry, came to the conclusion that the resistance of the cornea to deformation is negligibly small. The corneal bending forces involved in ...
متن کاملTheory and calibration of the Schiotz tonometer. I. The boundary surface of the cornea.
T JLo -onography as developed by Grant has been useful in both laboratory and clinic in elucidating the mechanisms of aqueous dynamics and of glaucoma and in assessing drug action. Tonography is the direct outgrowth of Friedenwald's" work on calibration of the Schi0tz tonometer. The present series of papers is devoted to a re-examination of some of the basic concepts of indentation tonometry in...
متن کاملRe-evaluation of the Schiotz tonometer calibration.
The calibration scale for the Schi0tz tonometers toas tested by comparing Goldmann applanation and Schi0tz measurements on the eyes of 906 patients. The study differed from several previous studies in that both the applanation and Schi0tz measurements were made with the subjects lying on their backs, thus eliminating errors due to change in intraocular pressure as the subject changed from one p...
متن کاملApplanation and Schi0tz tonometer standardizations for the owl monkey eye with a new technique for measuring episcleral venous pressure
A Schi0tz Pi calibration by open manometer was performed on 18 owl monkey eyes cannulated in situ. Pi values for the owl monkey corresponded closely to those of the 1955 Friedenwald nomogram. The average ocular rigidity measured in 16 additional eyes with the use of clinical techniques was 0.0173 (± 0.0015). A manometric calibration for a 4 mm. diameter applanation face was done similarly on 21...
متن کاملPressure measurement: which tonometer?
In the search for more accurate ways to determine intraocular pressure, the Halberg applanation tonometer may represent an alternative to the Schi0tz tonometer for routine screening pressure measurements. Although between 100 to 150 measurements are generally required to gain proficiency with the Halberg, the instrument is accurate in patients with normal corneas even with very abnormal scleral...