A novel radiation protection drape reduces radiation exposure during fluoroscopy guided electrophysiology procedures.

نویسندگان

  • Joseph J Germano
  • Gina Day
  • David Gregorious
  • Venkataraman Natarajan
  • Todd Cohen
چکیده

OBJECTIVE The purpose of this study was to evaluate a novel disposable lead-free radiation protection drape for decreasing radiation scatter during electrophysiology procedures. BACKGROUND In recent years, there has been an exponential increase in the number of electrophysiology (EP) procedures exposing patients, operators and laboratory staff to higher radiation doses. METHODS The RADPAD was positioned slightly lateral to the incision site for pectoral device implants and superior to the femoral vein during electrophysiology studies. Each patient served as their own control and dosimetric measurements were obtained at the examiner's elbow and hand. Radiation badge readings for the operator were obtained three months prior to RADPAD use and three months after introduction. RESULTS Radiation dosimetry was obtained in twenty patients: 7 electrophysiology studies, 6 pacemakers, 5 catheter ablations, and 2 implantable cardioverter-defibrillators. Eleven women and nine men with a mean age of 63 +/- 4 years had an average fluoroscopy time of 2.5 +/- 0.42 minutes per case. Mean dosimetric measurements at the hand were reduced from 141.38 +/- 24.67 to 48.63 +/- 9.02 milliroentgen (mR) per hour using the protective drape (63% reduction; p < 0.0001). Measurements at the elbow were reduced from 78.78 +/- 7.95 mR per hour to 34.50 +/- 4.18 mR per hour using the drape (55% reduction; p < 0.0001). Badge readings for three months prior to drape introduction averaged 2.45 mR per procedure versus 1.54 mR per procedure for 3 months post-initiation (37% reduction). CONCLUSION The use of a novel radiation protection surgical drape can significantly reduce scatter radiation exposure to staff and operators during a variety of EP procedures.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Evaluation of Patients’ Exposure during Angiography and Angioplasty Procedures in the Angiography Department of Shahid Madani Hospital in Tabriz

Introduction: Coronary angiography and angioplasty procedures lead to significant radiation exposure of patients. In the current study, the average radiation dose to patients during angiography in the Angiography Department of Shahid-Madani Hospital was determined. Materials and Methods: An image intensifier based angiography unit (Philips BH 3000) was used for angiography procedures. The accur...

متن کامل

Radiation protection education in fluoroscopy.

Potential biological damage from radiation received during fluoroscopy procedures is of particular concern because of the high volume and variety of procedures performed and the increasing length of radiation exposure. This article focuses on the effects of low-level radiation, gaps in education and skills among personnel performing and assisting with fluoroscopy, the certification and privileg...

متن کامل

Radiation Dose Reduction during Radial Cardiac Catheterization: Evaluation of a Dedicated Radial Angiography Absorption Shielding Drape

Objectives. Radiation scatter protection shield drapes have been designed with the goal of decreasing radiation dose to the operators during transfemoral catheterization. We sought to investigate the impact on operator radiation exposure of various shielding drapes specifically designed for the radial approach. Background. Radial access for cardiac catheterization has increased due to improved ...

متن کامل

Radiographer Delivered Fluoroscopy Reduces Radiation Exposure During Endoscopic Urological Procedures

INTRODUCTION The 1999 Ionising Radiation Regulations recommend that medical professionals using ionising radiation should aim to keep exposure as 'low as reasonably practicable'. Urologists regularly use fluoroscopy during endoscopic surgical procedures. In some institutions, this is delivered by a radiographer whereas in others, it is delivered by the urological surgeon. OBJECTIVES To determ...

متن کامل

Robotically driven interventions: a method of using CT fluoroscopy without radiation exposure to the physician.

Radiation exposure to physicians' hands during interventional procedures with computed tomography (CT) fluoroscopic guidance may be high. A robot was developed that could hold, orient, and advance a needle, with CT fluoroscopic guidance. This robot could be either computer or joystick controlled. Twenty-three robotically guided percutaneous interventions were performed without complication. Phy...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • The Journal of invasive cardiology

دوره 17 9  شماره 

صفحات  -

تاریخ انتشار 2005