Measurement of Cardiac Index by Transpulmonary Thermodilution Using an Implanted Central Venous Access Port: A Prospective Study in Patients Scheduled for Oncologic High-Risk Surgery
نویسندگان
چکیده
BACKGROUND Transpulmonary thermodilution allows the measurement of cardiac index for high risk surgical patients. Oncologic patients often have a central venous access (port-a-catheter) for chronic treatment. The validity of the measurement by a port-a-catheter of the absolute cardiac index and the detection of changes in cardiac index induced by fluid challenge are unknown. METHODS We conducted a monocentric prospective study. 27 patients were enrolled. 250 ml colloid volume expansions for fluid challenge were performed during ovarian cytoreductive surgery. The volume expansion-induced changes in cardiac index measured by transpulmonary thermodilution by a central venous access (CIcvc) and by a port-a-catheter (CIport) were recorded. RESULTS 23 patients were analyzed with 123 pairs of measurements. Using a Bland and Altman for repeated measurements, the bias (lower and upper limits of agreement) between CIport and CIcvc was 0.14 (-0.59 to 0.88) L/min/m2. The percentage error was 22%. The concordance between the changes in CIport and CIcvc observed during volume expansion was 92% with an r = 0.7 (with exclusion zone). No complications (included sepsis) were observed during the follow up period. CONCLUSIONS The transpulmonary thermodilution by a port-a-catheter is reliable for absolute values estimation of cardiac index and for measurement of the variation after fluid challenge. TRIAL REGISTRATION clinicaltrials.gov NCT02063009.
منابع مشابه
Pulmonary vascular permeability index in patients with femoral venous access for transpulmonary thermodilution: a prospective observational study
Title:Pulmonary vascular permeability index and global end-diastolic volume: are the data consistent in patients with femoral venous access for transpulmonary thermodilution: a prospective observational study
متن کاملA Comparison of Third-Generation Semi-Invasive Arterial Waveform Analysis with Thermodilution in Patients Undergoing Coronary Surgery
Uncalibrated semi-invasive continous monitoring of cardiac index (CI) has recently gained increasing interest. The aim of the present study was to compare the accuracy of CI determination based on arterial waveform analysis with transpulmonary thermodilution. Fifty patients scheduled for elective coronary surgery were studied after induction of anaesthesia and before and after cardiopulmonary b...
متن کاملTranspulmonary thermodilution using femoral indicator injection: a prospective trial in patients with a femoral and a jugular central venous catheter
INTRODUCTION Advanced hemodynamic monitoring using transpulmonary thermodilution (TPTD) is established for measurement of cardiac index (CI), global end-diastolic volume index (GEDVI) and extra-vascular lung water index (EVLWI). TPTD requires indicator injection via a central venous catheter (usually placed via the jugular or subclavian vein). However, superior vena cava access is often not fea...
متن کاملEffectiveness of Permanent Implantable Catheter (Polysite) in Children with Cancer
Abstract Background Totally implantable central venous access devices (ports) have been available for over 10 years, but have not been achieved widespread use in pediatric oncology patients. Ports facilitate the administration of chemotherapy in children with cancer. Materials and Methods In this study, early complications of implantable central venous access devices in children with diff...
متن کاملHow to Care for Implanted Ports
Implantable ports are used for intravenous infusion therapy and play an important role in management of oncology patients. These ports are best suited for patients requiring long-term therapy (>4 weeks). Implanted ports provide reliable venous access protect peripheral access increase patients’ comfort through reducing repeated and difficult vein punctures allow for safe and comfortable admi...