Suggesting a new design for multileaf collimator leaves based on Monte Carlo simulation of two commercial systems
نویسندگان
چکیده
Due to intensive use of multileaf collimators (MLCs) in clinics, finding an optimum design for the leaves becomes essential. There are several studies which deal with comparison of MLC systems, but there is no article with a focus on offering an optimum design using accurate methods like Monte Carlo. In this study, we describe some characteristics of MLC systems including the leaf tip transmission, beam hardening, leakage radiation and penumbra width for Varian and Elekta 80-leaf MLCs using MCNP4C code. The complex geometry of leaves in these two common MLC systems was simulated. It was assumed that all of the MLC systems were mounted on a Varian accelerator and with a similar thickness as Varian's and the same distance from the source. Considering the obtained results from Varian and Elekta leaf designs, an optimum design was suggested combining the advantages of three common MLC systems and the simulation results of this proposed one were compared with the Varian and the Elekta. The leakage from suggested design is 29.7% and 31.5% of the Varian and Elekta MLCs. In addition, other calculated parameters of the proposed MLC leaf design were better than those two commercial ones. Although it shows a wider penumbra in comparison with Varian and Elekta MLCs, taking into account the curved motion path of the leaves, providing a double focusing design will solve the problem. The suggested leaf design is a combination of advantages from three common vendors (Varian, Elekta and Siemens) which can show better results than each one. Using the results of this theoretical study may bring about superior practical outcomes.
منابع مشابه
Design, Simulation and Manufacture of a Multileaf Collimator to Confirm the Target Volumes in Intensity Modulated Radiation Therapy
Introduction: Intensity modulated radiation therapy (IMRT) is one of the cancer treatment methods. It is important to selectively aim at the target in this way, which can be performed using a multileaf collimator (MLC). Materials and Methods: In order to specifically irradiate the target volume in radiotherapy to reduce the patient absorbed dose, the use of multileaf collimator has been investi...
متن کاملMonte Carlo Study of Unflattened Photon Beams Shaped by Multileaf Collimator
Introduction: This study investigates basic dosimetric properties of unflattened 6 MV photon beam shaped by multileaf collimator and compares them with those of flattened beams.Materials and Methods: Monte Carlo simulation model using BEAM code was developed for a 6MV photon beam based on Varian Clinic 600 unique performance linac operated with and without a flattening filter in beam line. Dosi...
متن کاملInvestigation of Collimator Influential Parameter on SPECT Image Quality: a Monte Carlo Study
Background: Obtaining high quality images in Single Photon Emission Tomography (SPECT) device  is the most important goal in nuclear medicine. Because if image quality is low, the possibility of making a mistake in diagnosing and treating the patient will rise. Studying effective factors in spatial resolution of imaging systems is thus deemed to be vital. One of the most important factors in S...
متن کاملDerivation and validation of a sensitivity formula for knife-edge slit gamma camera: A theoretical and Monte Carlo simulation study
Introduction: Gamma cameras are proposed for online range verification and treatment monitoring in proton therapy. An Analytical formula was derived and validated for sensitivity of a slit collimator based on the photon fluence concept. Methods: Fluence formulation was generalized for photons distribution function and solved for high-energy point sources. The...
متن کاملDesign and Fabrication of the Mechanical Part of a Prototype Multileaf Collimator System
Introduction: Multileaf collimator (MLC) is state of the art among the radiation field shaping systems for conformal radiotherapy techniques. Some advantages of this system are as follow: reduction of treatment time, higher precision in conforming the radiation field to the projection of the target volume and lowering the chance of errors made by technologists. The purpose of this research was ...
متن کامل