An Intelligent Hand-held Microsurgical Instrument for Improved Accuracy

نویسندگان

  • Wei Tech Ang
  • Pradeep K. Khosla
  • Cameron N. Riviere
چکیده

This paper presents the development and initial experimental results of the first prototype of Micron, an active hand-held instrument to sense and compensate physiological tremor and other unwanted movement during vitreoretinal microsurgery. The instrument incorporates six inertial sensors, allowing the motion of the tip to be computed. The motion captured is processed to discriminate between desired and undesired components of motion. Tremor canceling is implemented via the weighted-frequency Fourier linear combiner (WFLC) algorithm, and compensation of non-tremorous error via a neural-network technique is being investigated. The instrument tip is attached to a three-degree-of-freedom parallel manipulator with piezoelectric actuation. The actuators move the tool tip in opposition to the tremor, thereby suppressing the erroneous motion. Motion canceling experiments with oscillatory motions in the frequency band of physiological tremor show that Micron is able to reduce error amplitude by 45.3% in 1-D tests and 34.3% in 3-D tests.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Design of all-accelerometer inertial measurement unit for tremor sensing in hand-held microsurgical instrument

We present the design of an all-accelerometer inertial measurement unit (IMU). The IMU forms part of an intelligent hand-held microsurgical instrument that senses its own motion, distinguishes between hand tremor and intended motion, and compensates in real-time the erroneous motion. The new IMU design consists of three miniature dual-axis accelerometers, two of which are housed in a sensor sui...

متن کامل

An Active Hand-Held Instrument for Enhanced Microsurgical Accuracy

This paper presents the first prototype of an active hand-held instrument to sense and compensate physiological tremor and other unwanted movement during vitreoretinal microsurgery. The instrument incorporates six inertial sensors (three accelerometers and three rate gyros) to detect motion of the handle. The movement of the instrument tip in three dimensions is then obtained using appropriate ...

متن کامل

A Study of Instrument Motion in Retinal Microsurgery

This paper reports on high-precision recordings of hand-held instrument motion during actual vitreoretinal microsurgery. The movement of a hand-held instrument during vitreoretinal microsurgery was recorded in six degrees of freedom. Data were acquired for 5 min using an inertial sensing module that has been developed for use with a commercially available microsurgical instrument. Maximum veloc...

متن کامل

Active handheld instrument for error compensation in microsurgery

Physiological hand tremor and other manual positioning errors limit precision in microsurgical procedures. Our research has involved development of adaptive algorithms and neural network methods for real-time compensation of such errors. This paper presents a novel design for an active hand-held microsurgical instrument to implement these algorithms, particularly during vitreoretinal microsurge...