Three-axis MR-conditional robot for high-intensity focused ultrasound for treating prostate diseases transrectally
نویسندگان
چکیده
BACKGROUND A prototype magnetic resonance image (MRI)-conditional robot was developed for navigating a high-intensity focused ultrasound (HIFU) system in order to treat prostate cancer transrectally. MATERIALS AND METHODS The developed robotic device utilizes three PC-controlled axes: a linear axis for motion along the rectum, an angular axis for rotation in the rectum, and a linear axis to lift the robot up and down. Experiments with the system were performed in a 1.5-T MRI system using gel phantoms. RESULT The robot was successfully operated in a 1.5-T clinical MRI system. The effect of piezoelectric motors and optical encoders was quantified based on the reduction of signal to noise ratio. Discrete and overlapping lesions were created accurately by moving the HIFU transducer with the robotic device. CONCLUSION An MRI-conditional HIFU robot was developed which can create controlled thermal lesions under MRI guidance. The intention is to use this robot transrectally in the future for the treatment of prostate cancer.
منابع مشابه
Phase I/II trial of high intensity focused ultrasound for the treatment of previously untreated localized prostate cancer.
PURPOSE We examined the safety and potential efficacy of transrectally delivered high intensity focused ultrasound for the full gland ablation of previously untreated localized prostate cancer. MATERIALS AND METHODS A total of 20 patients with localized prostate cancer underwent 1 to 3 high intensity focused ultrasound treatments of the prostate. The primary outcome was safety and the seconda...
متن کاملChronic effect of transrectal split-focus ultrasonic ablation on canine prostatic tissue.
The treatment time needed for high-intensity focused ultrasound (HIFU) ablation might be decreased substantially by using the split-focus approach, so we made a prototype 4.2-MHz split-focus therapeutic transducer combined with a small 6.5-MHz imaging ultrasonic probe for transrectally treatment of canine prostatic cancer and used it to experimentally evaluate the feasibility of using split-foc...
متن کاملDynamic Contrast-Enhanced MR Imaging in Detecting Local Tumor Progression after HIFU Ablation of Localized Prostate Cancer
Radical prostatectomy remains the standard treatment for patients with localized prostate cancer and a life expectancy exceeding 10 years (1). For patients with a life expectancy of less than 10 years who are not fit for surgery or do not want to experience the potential side effects of surgery, several alternative treatments have been introduced, such as, three-dimensional radiotherapy, brachy...
متن کاملThe road to clinical use of high-intensity focused ultrasound for liver cancer: technical and clinical consensus
Clinical use of high-intensity focused ultrasound (HIFU) under ultrasound or MR guidance as a non-invasive method for treating tumors is rapidly increasing. Tens of thousands of patients have been treated for uterine fibroid, benign prostate hyperplasia, bone metastases, or prostate cancer. Despite the methods' clinical potential, the liver is a particularly challenging organ for HIFU treatment...
متن کاملMR-ARFI and SWI to Detect Calcifications in the Brain in MRgHIFU Treatments
Introduction Calcifications in the brain can be associated with several intracranial pathologies including tumor growth, epilepsy, hyperthyriodism, and infectious diseases [1][2]. Recent advances have shown that MR guided High Intensity Focused Ultrasound (MRgHIFU) can be used in the brain, as a non-invasive functional neurosurgery [3]. Calcifications exhibit a high acoustic attenuation coeffic...