High SNR Acquisitions Improve the Repeatability of Liver Fat Quantification Using Confounder-corrected Chemical Shift-encoded MR Imaging

نویسندگان

  • Utaroh Motosugi
  • Diego Hernando
  • Curtis Wiens
  • Peter Bannas
  • Scott. B Reeder
چکیده

PURPOSE To determine whether high signal-to-noise ratio (SNR) acquisitions improve the repeatability of liver proton density fat fraction (PDFF) measurements using confounder-corrected chemical shift-encoded magnetic resonance (MR) imaging (CSE-MRI). MATERIALS AND METHODS Eleven fat-water phantoms were scanned with 8 different protocols with varying SNR. After repositioning the phantoms, the same scans were repeated to evaluate the test-retest repeatability. Next, an in vivo study was performed with 20 volunteers and 28 patients scheduled for liver magnetic resonance imaging (MRI). Two CSE-MRI protocols with standard- and high-SNR were repeated to assess test-retest repeatability. MR spectroscopy (MRS)-based PDFF was acquired as a standard of reference. The standard deviation (SD) of the difference (Δ) of PDFF measured in the two repeated scans was defined to ascertain repeatability. The correlation between PDFF of CSE-MRI and MRS was calculated to assess accuracy. The SD of Δ and correlation coefficients of the two protocols (standard- and high-SNR) were compared using F-test and t-test, respectively. Two reconstruction algorithms (complex-based and magnitude-based) were used for both the phantom and in vivo experiments. RESULTS The phantom study demonstrated that higher SNR improved the repeatability for both complex- and magnitude-based reconstruction. Similarly, the in vivo study demonstrated that the repeatability of the high-SNR protocol (SD of Δ = 0.53 for complex- and = 0.85 for magnitude-based fit) was significantly higher than using the standard-SNR protocol (0.77 for complex, P < 0.001; and 0.94 for magnitude-based fit, P = 0.003). No significant difference was observed in the accuracy between standard- and high-SNR protocols. CONCLUSION Higher SNR improves the repeatability of fat quantification using confounder-corrected CSE-MRI.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Reproducibility of Intra- and Inter-scanner Measurements of Liver Fat Using Complex Confounder-corrected Chemical Shift Encoded MRI at 3.0 Tesla

The purpose of this study was to prospectively evaluate the reproducibility of the proton density fat-fraction (PDFF) of the liver using the IDEAL algorithm, a quantitative confounder-corrected chemical-shift-encoded MRI method. Data were obtained from 15 volunteers on four different days. The first and the third examinations were conducted on scanner one with one-week intervals, while the seco...

متن کامل

Magnetic Resonance Imaging of Changes in Abdominal Compartments in Obese Diabetics during a Low-Calorie Weight-Loss Program

OBJECTIVES To investigate changes in the fat content of abdominal compartments and muscle area during weight loss using confounder-adjusted chemical-shift-encoded magnetic resonance imaging (MRI) in overweight diabetics. METHODS Twenty-nine obese diabetics (10/19 men/women, median age: 59.0 years, median body mass index (BMI): 34.0 kg/m2) prospectively joined a standardized 15-week weight-los...

متن کامل

Fast Single Breath-hold 3D Abdominal Spiral Imaging with Water / Fat Separation and Off-resonance Correction

Introduction Large volume coverage, short total scan time and robust water/fat separation are important issues in abdominal MRI. Breath-hold approaches are often used to avoid respiratory motion artifacts, which demands for efficient sampling schemes. Spiral imaging is one of them [1], but shows high sensitivity to off-resonance [2]. Image blurring resulting from spatially varying Bo inhomogene...

متن کامل

Spiral water-fat imaging with integrated off-resonance correction on a clinical scanner.

PURPOSE To integrate water-fat-resolved spiral gradient-echo imaging with off-resonance correction into a clinical MR scanner and to evaluate its basic feasibility and performance. MATERIALS AND METHODS Three-point chemical shift imaging was implemented with forward and strongly T(2)*-weighted reverse spiral sampling and with off-resonance correction after water-fat separation. It was applied...

متن کامل

MR validation and quantification of TACE treatment through direct visualization of Ethiodol with Chemical Shift Based Water-Fat Imaging

Figure 2: Images from a 71-year-old male with cirrhosis and a known HCC in the right lobe of the liver treated with ethiodol based TACE. An arrow shows the tumor location in each image. A) CT demonstrates a hyperdense region corresponding to the ethiodol distribution. Separated b) water, c) fat and d) calculated fat-fraction images are also shown. MR validation and quantification of TACE treatm...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 16  شماره 

صفحات  -

تاریخ انتشار 2017