In vivo verification of regional hyperthermia in the liver
نویسندگان
چکیده
PURPOSE We performed invasive thermometry to verify the elevation of local temperature in the liver during hyperthermia. MATERIALS AND METHODS Three 40-kg pigs were used for the experiments. Under general anesthesia with ultrasonography guidance, two glass fiber-optic sensors were placed in the liver, and one was placed in the peritoneal cavity in front of the liver. Another sensor was placed on the skin surface to assess superficial cooling. Six sessions of hyperthermia were delivered using the Celsius TCS electro-hyperthermia system. The energy delivered was increased from 240 kJ to 507 kJ during the 60-minute sessions. The inter-session cooling periods were at least 30 minutes. The temperature was recorded every 5 minutes by the four sensors during hyperthermia, and the increased temperatures recorded during the consecutive sessions were analyzed. RESULTS As the animals were anesthetized, the baseline temperature at the start of each session decreased by 1.3℃ to 2.8℃ (median, 2.1℃). The mean increases in temperature measured by the intrahepatic sensors were 2.42℃ (95% confidence interval [CI], 1.70-3.13) and 2.67℃ (95% CI, 2.05-3.28) during the fifth and sixth sessions, respectively. The corresponding values for the intraperitoneal sensor were 2.10℃ (95% CI, 0.71-3.49) and 2.87℃ (1.13-4.43), respectively. Conversely, the skin temperature was not increased but rather decreased according to application of the cooling system. CONCLUSION We observed mean 2.67℃ and 2.87℃ increases in temperature at the liver and peritoneal cavity, respectively, during hyperthermia. In vivo real-time thermometry is useful for directly measuring internal temperature during hyperthermia.
منابع مشابه
In vivo dose verification using using an amorphous silicon flat panel-type imager (a-Si EPIDs)
Introduction: Electronic portal imaging devices (EPIDs) could be used to dose verification of radiotherapy treatment plans. In vivo dose verification is performed to reduce differences found between dose delivered to the patient and the prescribed dose. The aim of this study was to perform a fast and efficient technique for the verification of delivered dose to the patient usin...
متن کاملAssessment of a 2D EPID-based Dosimetry Algorithm for Pre-treatment and In-vivo Midplane Dose Verification
Introduction: The use of electronic portal imaging devices (EPIDs) is a method for the dosimetric verification of radiotherapy plans both pretreatment and in-vivo. The aim of this study was to test a 2D EPID-based dosimetry algorithm for dose verification of some plans inside a homogenous and anthropomorphic phantom and in-vivo, as well. Materials and Methods: </strong...
متن کاملSimulation of tissue heating by magnetic fluid hyperthermia
Objective: Magnetic fluid hyperthermia is a technique in which thermal energy is generated by magnetic nanoparticles (MNPs) that are excited by an alternating magnetic field (AC field). During hyperthermia, in-vivo monitoring of elevation of temperature relies on invasive insertion of conventional thermometers, or employment of thermo-sensitive cameras that lack high precision....
متن کاملVerification of the Accuracy of the Delivered Dose in Brain Tumors by in Vivo Dosimetry Using Diode Detectors
Introduction: During radiotherapy, high accuracy in the dose delivery is required because there is a strong relationship between the absorbed dose, local tumor control and particularly the normal tissue damage. In many institutions, in vivo dosimetry using diodes is performed to check the actual dose delivered. In general, the uncertainty in the dose delivered should fall within ± 5% of the...
متن کاملEffect of the combination of 6 MeV radiotherapy with hyperthermia and gold nanoparticles on the MCF-7 breast cancer cells
Introduction: Combining radiotherapy as one of the main modalities used for cancer treatment with other modalities such as hyperthermia has recently played a special role in reducing side effects and improving treatment outcomes. In addition, Gold nanoparticles (GNPs) have also attracted attention as suitable clinical agents for enhancing the effect of radiotherapy in treating ...
متن کامل