A New Approach for Scatter Removal and Attenuation Compensation from SPECT/CT Images
نویسندگان
چکیده
OBJECTIVE(S) In SPECT, the sinogram contains scatter and lack of attenuated counts that degrade the reconstructed image quality and quantity. Many techniques for attenuation and scatter correction have been proposed. An acceptable method of correction is to incorporate effects into an iterative statistical reconstruction. Here, we propose new Maximum Likelihood Expectation Maximization (MLEM) formula to correct scattering and attenuating photons during reconstruction. MATERIALS AND METHODS In this work, scatters are estimated through Klein-Nishina formula in all iterations and CT images are used for accurate attenuation correction. Reconstructed images resulted from different MLEM reconstruction formula have been compared considering profile agreement, contrast, mean square error, signal-to-noise ratio, contrast-to-noise ratio and computational time. RESULTS The proposed formula has a good profile agreement, increased contrast, signal-to-noise (SNR) & contrast-to-noise ratio (CNR), computational time and decreased mean square error (MSE) compared with uncorrected images and/or images from conventional formula. CONCLUSION In conclusion, by applying the proposed formula we were able to correct attenuation and scatter via MLEM and improve the image quality, which is a necessary step for both qualitative and quantitative SPECT images.
منابع مشابه
A New Approach for Scatter Removal and Attenuation Compensation from SPECT/CT Images
Objective(s): In SPECT, the sinogram contains scatter and lack of attenuated counts that degrade the reconstructed image quality and quantity. Many techniques for attenuation and scatter correction have been proposed. An acceptable method of correction is to incorporate effects into an iterative statistical reconstruction. Here, we propose new Maximum Likelihood Expectation Maximiz...
متن کاملA new approach to scatter correction in SPECT images based on Klein_Nishina equation
Introduction: Scattered photon is one of the main defects that degrade the quality and quantitative accuracy of nuclear medicine images. Accurate estimation of scatter in projection data of SPECT is computationally extremely demanding for activity distribution in uniform and non-uniform dense media. Methods: The objective of this paper is to develop and validate a scatter correction technique ...
متن کاملNew approach for attenuation correction in SPECT images, using linear optimization
Background: Photon attenuation as an inevitable physical phenomenon influences on the diagnostic information of SPECT images and results to errors in accuracy of quantitative measurements. This can be corrected via different physical or mathematical approaches. As the correction equation in mathematical approaches is nonlinear, in this study a new method of linearization called ‘Piece ...
متن کاملA New Approach for Quantitative Evaluation of Reconstruction Algorithms in SPECT
ABTRACT Background: In nuclear medicine, phantoms are mainly used to evaluate the overall performance of the imaging systems and practically there is no phantom exclusively designed for the evaluation of the software performance. In this study the Hoffman brain phantom was used for quantitative evaluation of reconstruction techniques. The phantom is modified to acquire t...
متن کاملQuantitative SPECT and planar 32P bremsstrahlung imaging for dosimetry purpose –An experimental phantom study
Background: In this study, Quantitative 32P bremsstrahlung planar and SPECT imaging and consequent dose assessment were carried out as a comprehensive phantom study to define an appropriate method for accurate Dosimetry in clinical practice. Materials and Methods: CT, planar and SPECT bremsstrahlung images of Jaszczak phantom containing a known activity of 32P were acquired. In addition, Phanto...
متن کامل