Survey on Prevacuum High-pressure Steam Sterilizers.

نویسندگان

  • E M DARMADY
  • S E DREWETT
  • K E HUGHES
چکیده

None of the 10 prevacuum high-pressure sterilizers of different makes tested was able to produce and maintain the conditions advocated by the Medical Research Council working party on high-pressure steam sterilizers (1959) or by Knox and Penikett (1958) with the result that steam did not penetrate adequately the single challenge load and it was not sterilized. The sterilization of ;group drums' of various sizes and contents was erratic and tended to give operators a false sense of security. An alarming number of minor engineering faults were present in seven out of 10 machines tested and they require very much more skilled maintenance than is being given at the moment. The possibility of centralizing sterilizers to central sterile supply departments and placing them under the care of a regional engineer cannot be too highly recommended. The presence of undetected ;leaks' and a failure to draw a prevacuum of 20 mm. even with a steam burst interferes with sterilization of a challenge load. A leak test should be performed twice daily and should not exceed more than 1 mm. in one minute at 20 mm. absolute. The centre of the load should be monitored by crossed tapes or Brownes tubes in each sterilizing cycle. Although the challenge load was sterilized when the chamber was filled to capacity, a more reliable cycle consisting of a double prevacuum of 20 mm. or more with intermediate steam burst to 10 lb. ensured the sterilizing of a single challenge load, which could be adequately controlled by the chamber drain temperature.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Monitoring sterilization of dressings in high-vacuum pressure-steam sterilizers.

Undetected air leaks in a high-vacuum sterilizer can defeat the sterilization process. For this reason the maintenance and supervision of these sterilizers needs expert engineering care. The importance of controlling the high-vacuum sterilization process from the centre of a load rather than the chamber drain is emphasized, and a method of detecting any failure in a high-vacuum process is descr...

متن کامل

The testing of sterilizers. 2. Thermophilic spore papers.

Satisfactory spore preparations for testing hospital sterilizers can be made from B. stearothermophilus and comparatively simple methods for loading and drying the papers are considered to be valid. Each batch of spore papers should be calibrated by plotting dose-response curves for exposure to steam. A minimum standard in terms of the L.D.50 to moist heat is proposed and confidence limits sugg...

متن کامل

steam-formaldehyde sterilizer for gas penetration efficiency

SYNOPSIS A test piece is described for monitoring the performance of low-temperature steam-withformaldehyde sterilizers. Comparative tests have shown it to be more difficult to penetrate than an arterial catheter when exposed to the same sterilizing conditions. It is permanent and simple to use and maintain. The growth or non-growth of bacterial spores, in the convenient form of spore strips, i...

متن کامل

Testing a steam-formaldehyde sterilizer for gas penetration efficiency.

A test piece is described for monitoring the performance of low-temperature steam-with-formaldehyde sterilizers. Comparative tests have shown it to be more difficult to penetrate than an arterial catheter when exposed to the same sterilizing conditions. It is permanent and simple to use and maintain. The growth or non-growth of bacterial spores, in the convenient form of spore strips, is used t...

متن کامل

تحلیل انرژی و اگزرژی کوپلینگ نیروگاه سیکل ترکیبی و آب‌شیرین‌کن تبخیری چندمرحله‌ای

Due to an increasing need of energy, lack of energy resources worldwide, it is very important to optimize energy generation plant to improve efficiency and production rate .One of the most important plant, for production of water and power are dual purpose power plants. In this kind of power plant a desalination system combines with power generation cycle. According to these issues, in this pap...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Journal of clinical pathology

دوره 17  شماره 

صفحات  -

تاریخ انتشار 1964