Externalized Conductor Cables in QuickSite Left Ventricular Pacing Lead and Riata Right Ventricular Lead in a Single Patient: A Common Problem With Silicone Insulation
نویسندگان
چکیده
QuickSite (St Jude Medical, Sylmar, CA, USA) is a silicone and polyurethane-insulated coronary sinus pacing lead. Riata lead (St Jude Medical, Sylmar, CA, USA) is a silicone insulated right ventricular shock lead. Recently, insulation breach of silicone based leads raised a huge concern. Fluoroscopic examination of these two leads in the same patient revealed externalization of these two leads. Same mechanism producing insulation breach of Riata lead may be involved in externalization of QuickSite LV lead as distal part of insulation is also made of silicone.
منابع مشابه
Shock-induced right ventricular pacing failure caused by a short circuit: Uncommon but life-threatening complication of the Riata lead
Introduction The Riata and Riata ST implantable cardioverter-defibrillator (ICD) leads, a narrow caliber silicone lead without an outer jacket, have been shown to be prone to structural and electrical failure. While much interest has been shown in potentially lethal high-voltage (HV) failures caused by electrical short circuits due to a breach in the ethylene tetrafluoroethylene (ETFE) insulati...
متن کاملFluoroscopic investigation of Riata® transvenous defibrillator leads.
BACKGROUND Implantable cardioverter-defibrillator leads from Riata® family (St. Jude Medical Inc., Sylmar, CA, USA) have been recently recalled by Food and Drug Administration for concerns of a unique type of "inside-out" insulation failure leading to conductor externalization. The objective of this study was to evaluate the prevalence and predictors of conductor externalization in patients imp...
متن کاملPrevalence and presentation of externalized conductors and electrical abnormalities in Riata defibrillator leads after fluoroscopic screening: report from the Netherlands Heart Rhythm Association Device Advisory Committee.
BACKGROUND The Riata family of implantable cardioverter-defibrillator (ICD) leads is prone to a specific insulation abrasion characterized by externalization of conductor cables. The objective of this study was to determine the prevalence of externalized conductors and electrical abnormalities in Riata ICD leads by fluoroscopic screening and standard ICD interrogation. METHODS AND RESULTS All...
متن کاملRepetitive inappropriate implantable cardioverter-defibrillator shocks due to insulation failure with externalized conductor cables of a Biotronik Linox SD ICD lead.
We report the case of a 54-year-old patient with repetitive inappropriate implantable cardioverter-defibrillator (ICD) shocks due to externalized pace-sense conductor cables of a Biotronik Linox SD 65/16 ICD lead. This case report is one of few in the literature describing inappropriate shocks due to insulation failure with externalization of conductor cables of an ICD lead from the Biotronik L...
متن کاملExternalized conductors and insulation failure in Biotronik defibrillator leads: History repeating or a false alarm?
Conductor externalization and insulation failure are frequent complications with the recalled St. Jude Medical Riata implantable cardioverter-defibrillator (ICD) leads. Conductor externalization is a "unique" failure mechanism: Cables externalize through the insulation ("inside-out" abrasion) and appear outside the lead body. Recently, single reports described a similar failure also for Biotron...