Study of scattered radiation during fluoroscopy in hip surgery*
نویسندگان
چکیده
OBJECTIVE To measure the scattered radiation dose at different positions simulating hip surgery. MATERIALS AND METHODS We simulated fluoroscopy-assisted hip surgery in order to study the distribution of scattered radiation in the operating room. To simulate the patient, we used a anthropomorphic whole-body phantom, and we used an X-ray-specific detector to quantify the radiation. Radiographs were obtained with a mobile C-arm X-ray system in continuous scan mode, with the tube at 0º (configuration 1) or 90º (configuration 2). The operating parameters employed (voltage, current, and exposure time) were determined by a statistical analysis based on the observation of orthopedic surgical procedures involving the hip. RESULTS For all measurements, higher exposures were observed in configuration 2. In the measurements obtained as a function of height, the maximum dose rates observed were 1.167 (± 0.023) µSv/s and 2.278 (± 0.023) µSv/s in configurations 1 and 2, respectively, corresponding to the chest level of health care professionals within the operating room. Proximal to the patient, the maximum values were recorded in the position occupied by the surgeon. CONCLUSION We can conclude that, in the scenario under study, health care professionals workers are exposed to low levels of radiation, and that those levels can be reduced through the use of personal protective equipment.
منابع مشابه
Estimation of the effective dose to the radiologists during fluoroscopy or angiography of abdominal viscera
ABSTRACTBackground: as fluoroscopy or different angiographies are usually evaluated using TL dosimeter. Thismethod is a time-consuming procedure therefore, radiologists are usually worried and unawareof their exact radiation dose received during each fluoroscopy or angiography. In this study, anew trend for quick estimation of effective dose based on measuring air-absorbed dose of thescattered ...
متن کاملبررسی میزان پرتو پراکنده Xو دوز دریافتی کارکنان بخش آنژیوگرافی یک بیمارستان نظامی
Background and aims: Using X-ray for diagnostic test, is one of the most important exposure to radiation sources for general population and in particular for radiation workers. Therefore, holding down the pollution at the lowest possible level is very important for protection against the internal exposure. The objective of this study is to survey the scattered X-ray rate and received dose by st...
متن کاملEvaluation of Patients’ Exposure during Angiography and Angioplasty Procedures in the Angiography Department of Shahid Madani Hospital in Tabriz
Introduction: Coronary angiography and angioplasty procedures lead to significant radiation exposure of patients. In the current study, the average radiation dose to patients during angiography in the Angiography Department of Shahid-Madani Hospital was determined. Materials and Methods: An image intensifier based angiography unit (Philips BH 3000) was used for angiography procedures. The accur...
متن کاملManagement of Hip Fractures in Lateral Position without a Fracture Table
Background: Hip fracture Management in supine position on a fracture table with biplane fluoroscopic views has some difficulties which leads to prolongation of surgery and increasing x- rays’ dosage. The purpose of this study was to report the results and complications of hip fracture management in lateral position on a conventional operating table with just anteroposterior fluoroscopic view. ...
متن کاملThe Relationship Between Body Mass Index and Fluoroscopy Time During Intraarticular Hip Injection: A Multicenter Cohort Study.
BACKGROUND Higher body mass index (BMI) is associated with difficulty in obtaining imaging studies. While there is a small body of literature regarding the relationship between fluoroscopy time and BMI during injections for pain management, this has not been studied for intraarticular (IA) hip injections. Further, in academic training centers, trainee involvement may affect this relationship. ...
متن کامل