An evaluation of tympanic thermometry in a paediatric emergency department.
نویسندگان
چکیده
BACKGROUND AND OBJECTIVES The consequences of missing fever in children can be grave. Body temperature is commonly recorded at the axilla but accuracy is a problem. This study aimed to evaluate the accuracy of a tympanic thermometer in the paediatric emergency setting. METHOD In a total of 106 infants, the body temperature was measured in the daytime with an infrared tympanic thermometer, and at the axilla with an electronic thermometer and at the rectum (gold standard for measurement of body temperature). Fever was defined as a rectal temperature of 38.0 degrees C or greater, axillary temperature of 37.5 degrees C or greater, and tympanic temperature of 37.6 degrees C or greater. The temperature readings at the three sites were compared statistically. RESULTS There was a greater correlation of the tympanic measurement with the rectal measurement than the axillary with the rectal in both febrile and afebrile infants. The mean difference between the tympanic and rectal measurements was lower than that between the axillary and rectal measurements in both groups of infants (tympanic 0.38 degrees C and 0.42 degrees C, and axillary 1.11 degrees C and 1.58 degrees C, respectively). Tympanic measurements had a sensitivity of 76% whereas axillary measurements had a sensitivity of only 24% with rectal temperatures of 38-38.9 degrees C. CONCLUSION Tympanic thermometry is more accurate than measurement of temperature with an electronic axillary thermometer. It is also quick and safe, and thus it is recommended in the paediatric emergency setting.
منابع مشابه
Accuracy of a noninvasive temporal artery thermometer for use in infants.
OBJECTIVES To assess the accuracy of a new noninvasive temporal artery (TA) thermometer in infants; to compare the accuracy of the TA thermometer with that of a tympanic thermometer, using rectal thermometry as the criterion standard; and to compare the tolerability of the TA thermometer with that of the tympanic and rectal thermometers. DESIGN Prospective evaluation of the accuracy of TA and...
متن کاملOral and Tympanic Membrane Temperatures Are Inaccurate to Identify Fever in Emergency Department Adults
INTRODUCTION Identifying fever can influence management of the emergency department (ED) patient, including diagnostic testing, treatment, and disposition. We set out to determine how well oral and tympanic membrane (TM) temperatures compared with rectal measurements. METHODS A convenience sample of consecutively adult ED patients had oral, TM, and rectal temperatures performed within several...
متن کاملBed Utilisation in an Irish Regional Paediatric Unit – A Cross-Sectional Study Using the Paediatric Appropriateness Evaluation Protocol (PAEP)
Background Increasing demand for limited healthcare resources raises questions about appropriate use of inpatient beds. In the first paediatric bed utilisation study at a regional university centre in Ireland, we conducted a cross-sectional study to audit the utilisation of inpatient beds at the Regional Paediatric Unit (RPU) in University Hospital Limerick (UHL), Limerick, Ireland and also exa...
متن کاملA comparison of tympanic and rectal temperatures in term NIGERIAN neonates
BACKGROUND Tympanic thermometry has come as a suitable alternative to traditional thermometry because of its safety and ease of use. However, it is still yet to gain wide acceptance in African settings due to conflicting results on its accuracy, thus rectal thermometry remains the gold standard in the newborn. The aim of this study was to compare tympanic and rectal temperatures in term Nigeria...
متن کاملThe use of infrared thermometry for the detection of fever.
The use of tympanic thermometry is attractive in primary care, but a recent systematic review highlighted the paucity of data comparing tympanic thermometry with conventional methods. We report a study of 94 preschool children presenting to primary care in the United Kingdom (UK) with acute cough in whom tympanic infrared and axillary mercury thermometry are compared. Infrared thermometry showe...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Emergency medicine journal : EMJ
دوره 23 1 شماره
صفحات -
تاریخ انتشار 2006