Clinical and Wear Analyses of 9 Large Metal-on-Metal Total Hip Prostheses
نویسندگان
چکیده
BACKGROUND Metal-on-Metal (MoM) total hip arthroplasties (THA) are associated with pseudotumor formation and high revision rates. This prospective study analysed the clinical and wear analyses of 9 large Metal-on-Metal (MoM) total hip arthroplasties (THA) to understand the underlying mechanisms of failure. The MoM bearings were revised for multiple reasons; the main reason was pseudotumor formation. MATERIALS AND METHODS From 2006 till 2010 the Reinier de Graaf Hospital implanted 160 large head M2a-Magnum™ (Biomet Inc. Warsaw, Indiana, USA) THAs in 150 patients. The first year, 9 bearings were revised and analysed at the Biomechanics Section, Hamburg University of Technology, Germany. We performed clinical (Harris Hip Score, radiographic analysis, blood cobalt and chromium) and wear analysis (implant, tissue and fluid) of the 9 bearings. Since this study did not fall under the scope of the Medical Research Involving Human Subjects Act in The Netherlands, no ethical approval was necessary. In this prospective study all patient details were anonymized by the corresponding author, all other authors were blinded during the research and wear analyses. Patients with bilateral MoM implants were excluded. RESULTS The 9 bearings had a median (IQR) survival of 41.0 (25) months in situ. From these bearings, three showed no noticeable wear. The median (IQR) head wear volume was 3.2 (3.6) mm3 and maximum wear depth 0.02 (0.02) mm. For the cup the median (IQR) wear volume was 0.23 (0.3) mm3 with a maximum wear depth of 0.03 (0.05) mm. CONCLUSION An early identification of parameters related to failure of the MoM THA, such as pain, decreased range of motion, radiographic changes and high levels of blood cobalt and chromium is of great importance for patient's quality of life. Especially now patients and surgeons face the long term effects of all these bearings still in situ. This study reports the clinical and wear analyses of 9 MoM THA. In the majority of this group the reason for revision was pseudotumor formation. Most bearings showed signs of wear, however with a great diversity in clinical analysis, in inclination angle, serum cobalt and chromium levels as well as wear analysis. For a better understanding of the underlying mechanisms related with failure, more wear analyses of revised MoM bearings are necessary as well as a frequent follow-up of the patients with a MoM bearing.
منابع مشابه
Influence of simulator kinematics on the wear of metal-on-metal hip prostheses.
There is now considerable interest in metal-on-metal bearings for hip prostheses. Extremely low wear rates (0.1 mm3/10(6) cycles) have been reported in some simulator studies, while in vivo studies, although still very low, have shown wear rates of the order of 1 mm3/10(6) cycles. The aim of this study was to compare wear rates of metal-on-metal bearings in two hip simulators with different kin...
متن کاملTaper junction failure in large-diameter metal-on-metal bearings
OBJECTIVES An ongoing prospective study to investigate failing metal-on-metal hip prostheses was commenced at our centre in 2008. We report on the results of the analysis of the first consecutive 126 failed mated total hip prostheses from a single manufacturer. METHODS Analysis was carried out using highly accurate coordinate measuring to calculate volumetric and linear rates of the articular...
متن کاملMetal release from hip prostheses: cobalt and chromium toxicity and the role of the clinical laboratory.
Individuals with advanced hip disease suffer from pain, impaired hip function, and decreased quality of life. Roughly one million metal-on-metal (MoM) hip prostheses have been implanted worldwide in order to ameliorate these issues. While most MoM hip replacements are successful, some patients suffer from serious adverse effects secondary to the release of metal debris due to implant wear and c...
متن کاملOne-year prospective comparative study of three large-diameter metal-on-metal total hip prostheses: serum metal ion levels and clinical outcomes.
INTRODUCTION The good clinical outcomes and low wear obtained with 28-mm metal-on-metal implants for total hip replacement prompted the development of large-diameter heads that more closely replicated the normal hip anatomy, with the goal of improving prosthesis stability. However, the blood release of metal ions due to wear at the bearing surfaces and the high rate of groin pain seen with larg...
متن کاملAnalysis of 118 second-generation metal-on-metal retrieved hip implants.
Osteolysis is due to particulate wear debris and is responsible for the long-term failure of total hip replacements. It has stimulated the development of alternative joint surfaces such as metal-on-metal or ceramic-on-ceramic implants. Since 1988 the second-generation metal-on-metal implant Metasul has been used in over 60 000 hips. Analysis of 118 retrieved specimens of the head or cup showed ...
متن کامل