Commissioning a 50–100 kV X‐ray unit for skin cancer treatment

نویسندگان

  • Ren‐Dih Sheu
  • Allison Powers
  • Yeh‐Chi Lo
چکیده

This study provides the authors' experience along with dosimetric data from the commissioning of two Sensus SRT-100 50-100 kV X-ray units. Data collected during the commissioning process included: a) HVL, b) output (dose rate), c) applicator cone factors, and d) percentage depth dose. A Farmer-type chamber (PTW-N23333), and a thin-window parallel plate ion chamber (PTW-N23342) were used for dose rate measurements and dose profiles were measured with EBT3 GafChromic film. The average HVL values for 50, 70, and 100 kV of the two treatment units were found to be 0.52, 1.15, and 2.20 mm Al, respectively. The HVL's were 5%-9% lower when measured with the Farmer chamber, as compared to measurements with the parallel plate chamber, for energies of 70 and 100 kV. Dose rates were also measured to be 3%-4% lower with the Farmer chamber. The dose rate variation between the two units was found to be 2%-9% for 50, 70, and 100 kV. The dose uniformity over a circle of 2 cm diameter was within 4% in four cardinal directions; however, the dose profiles for the 5 cm applicator were nonuniform, especially in the cathode-anode direction. Measurements indicated as much as 15% lower dose for the 50 kV beam at field edge on the anode side, when normalized to the center. The crossline profile was relatively more symmetric, with a maximum deviation of 10% at the field edge. All ion chamber readings agreed with film measurements within 3%. The nonuniform profile produced by these units may introduce uncertainty in dose rate measurements, especially for larger applicators. Since there is no intrinsic tool (crosshair or field light) for alignment with the beam axis, the user should take care when positioning the chamber for output measurements. The data obtained with a Farmer-type chamber should be used cautiously and as a reference only for the SRT-100 X-ray treatment unit.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Evaluation of a systemic change of Superficial X-ray tube characteristics

Introduction: Very low, Low and Medium X-ray machines are used extensively for Superficial Radiation Therapy mostly used for Basal Cell, Squamous Cell Carcinoma, and Melanoma skin cancers. In most of cases, Superficial X-ray Therapy (SXRT) is preferred to Electron beams generated using a linear accelerator. SXRT tube characteristics may vary with the use of the machine for a lo...

متن کامل

Dosimetric characteristics of a new unit for electronic skin brachytherapy

PURPOSE Brachytherapy with radioactive high dose rate (HDR) (192)Ir source is applied to small skin cancer lesions, using surface applicators, i.e. Leipzig or Valencia type. New developments in the field of radiotherapy for skin cancer include electronic brachytherapy. This technique involves the placement of an HDR X-ray source close to the skin, therefore combining the benefits of brachythera...

متن کامل

Determination of hemolysis, osmotic fragility and fluorescence anisotropy on irradiated red blood cells as a function of kV of medical diagnostic X-rays

Background: People occasionally undergo medical diagnostic X-ray examinations and expose their red blood cells to radiation. Radiation that is generated from medical diagnostic X-ray machines is widely used in medical diagnoses. One of the important parameters is kilo-voltage (kV) that is applied across the X-ray tube in medical diagnostic X-ray machines. Kilo-voltage influences the radiation d...

متن کامل

Calculation of Absorbed Glandular Dose using a FORTRAN Program Based on Monte Carlo X-ray Spectra in Mammography

Introduction: Average glandular dose calculation in mammography with Mo-Rh target-filter and dose calculation for different situations is accurate and fast. Material and Methods: In this research, first of all, x-ray spectra of a Mo target bombarded by a 28 keV electron beam with and without a Rh filter were calculated using the MCNP code. Then, we used the Sobol-Wu parameters to write a FORTRA...

متن کامل

Commissioning and quality assurances of the Intrabeam Intra-Operative radiotherapy unit

Purpose: The authors report comprehensive commissioning and quality assurance (QA) procedures for Intrabeam, Intra-Operative radiotherapy (IORT) unit. The Intrabeam system miniature X-ray source is a 50 kV and 40 μA unit. Methods: The authors’ tests include measurements of output, beam deflection, isotropy, kVp and mAs measurements, quality index, isodose, reproducibility, linearity, depth dose...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 16  شماره 

صفحات  -

تاریخ انتشار 2015