An organ and effective dose study of XVI and OBI cone‐beam CT systems

نویسندگان

  • Daniel E. Hyer
  • Christopher F. Serago
  • Siyong Kim
  • Jonathan G. Li
  • David E. Hintenlang
چکیده

The main purpose of this work was to quantify patient organ doses from the two kilovoltage cone beam computed tomography (CBCT) systems currently available on medical linear accelerators, namely the X-ray Volumetric Imager (XVI, Elekta Oncology Systems) and the On-Board Imager (OBI, Varian Medical Systems). Organ dose measurements were performed using a fiber-optic coupled (FOC) dosimetry system along with an adult male anthropomorphic phantom for three different clinically relevant scan sites: head, chest, and pelvis. The FOC dosimeter was previously characterized at diagnostic energies by Hyer et al. [Med Phys 2009;36(5):1711-16] and a total uncertainty of approximately 4% was found for in-phantom dose measurements. All scans were performed using current manufacturer-installed clinical protocols and appropriate bow-tie filters. A comparison of image quality between these manufacturer-installed protocols was also performed using a Catphan 440 image quality phantom. Results indicated that for the XVI, the dose to the lens of the eye (1.07 mGy) was highest in a head scan, thyroid dose (19.24 mGy) was highest in a chest scan, and gonad dose (29 mGy) was highest in a pelvis scan. For the OBI, brain dose (3.01 mGy) was highest in a head scan, breast dose (5.34 mGy) was highest in a chest scan, and gonad dose (34.61 mGy) was highest in a pelvis scan. Image quality measurements demonstrated that the OBI provided superior image quality for all protocols, with both better spatial resolution and low-contrast detectability. The measured organ doses were also used to calculate a reference male effective dose to allow further comparison of the two machines and imaging protocols. The head, chest, and pelvis scans yielded effective doses of 0.04, 7.15, and 3.73 mSv for the XVI, and 0.12, 1.82, and 4.34mSv for the OBI, respectively.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Evaluation of Factors Affecting Absorbed Dose, Optimization Methods and Patient Dose Reduction in Dental Cone Beam CT (CBCT)

Introduction: Nowadays the use of cone beam computed tomography in dental imaging is increasing, although this method has a much lower dose than conventional CT scans, it delivers a higher dose than the panoramic and periapical patients therefore, , the aim of this study was to investigate the factors affecting the patients' dose in dental CBCTs and methods of optimizing and reducing the patien...

متن کامل

Evaluation of the gray level in CBCT systems and its relationship with HU in CT Scanners

Introduction: Cone-beam CT (CBCT) is an imaging system which offers three-dimensional (3D), multiplanar images and has many advantages over computed tomography (CT) including shorter acquisition times for the resolution desired in dentistry, lower radiation dose to the patient, reasonable price and higher spatial resolution but CBCT scanners are unable to display actual Hounsf...

متن کامل

Automatic image registration performance for two different CBCT systems; variation with imaging dose

The performance of an automatic image registration algorithm was compared on image sets collected with two commercial CBCT systems, and the relationship with imaging dose was explored. CBCT images of a CIRS Virtually Human Male Pelvis phantom (VHMP) were collected on Varian TrueBeam/OBI and Elekta Synergy/XVI linear accelerators, across a range of mAs settings. Each CBCT image was registered 10...

متن کامل

The Impact of Residual Geometric Inaccuracies on Normal Organ Doses in Image Guided-Radiation Therapy of Prostate Cancer Using On-Board Kilovoltage Cone-Beam Computed Tomography

Introduction: The aim of this retrospective study was to evaluate the variations in delivered dose to the bladder, rectum, and femoral heads of prostate cancer patients during a course of treatment by image-guided radiation therapy (IGRT). Materials and Methods: Overall, 15 patients with prostate cancer were selected and. Each week, for each patient five consecutive cone beam computed tomograph...

متن کامل

Investigation of the radiation dose from cone‐beam CT for image‐guided radiotherapy: A comparison of methodologies

Four methodologies were evaluated for quantifying kilovoltage cone-beam computed tomography (CBCT) dose: the Cone-Beam Dose Index (CBDI), IAEA Report 5 recommended methodology (IAEA), the AAPM Task Group 111 methodology (TG111), and the current dose metric; the Computed Tomography Dose Index (CTDI) on two commercial Varian cone-beam CT imaging systems; the Clinac iX On-Board Imager (OBI); and t...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 11  شماره 

صفحات  -

تاریخ انتشار 2010