Is the international normalised ratio (INR) reliable? A trial of comparative measurements in hospital laboratory and primary care settings.
نویسندگان
چکیده
AIM To determine the reliability of international normalised ratio (INR) measurement in primary care by practice nurses using near patient testing (NPT), in comparison with results obtained within hospital laboratories by varied methods. METHODS As part of an MRC funded study into primary care oral anticoagulation management, INR measurements obtained in general practice were validated against values on the same samples obtained in hospital laboratories. A prospective comparative trial was undertaken between three hospital laboratories and nine general practices. All patients attending general practice based anticoagulant clinics had parallel INR estimations performed in general practice and in a hospital laboratory. RESULTS 405 tests were performed. Comparison between results obtained in the practices and those in the reference hospital laboratory (gold standard), which used the same method of testing for INR, showed a correlation coefficient of 0.96. Correlation coefficients comparing the results with the various standard laboratory techniques ranged from 0.86 to 0.92. It was estimated that up to 53% of tests would have resulted in clinically significant differences (change in warfarin dose) depending upon the site and method of testing. The practice derived results showed a positive bias ranging from 0.28 to 1.55, depending upon the site and method of testing. CONCLUSIONS No technical problems associated with INR testing within primary care were uncovered. Discrepant INR results are as problematic in hospital settings as they are in primary care. These data highlight the failings of the INR to standardise when different techniques and reagents are used, an issue which needs to be resolved. For primary care to become more involved in therapeutic oral anticoagulation monitoring, close links are needed between hospital laboratories and practices, particularly with regard to training and quality assurance.
منابع مشابه
SMART: Self-Management of Anticoagulation, a Randomised Trial [ISRCTN19313375]
BACKGROUND Oral anticoagulation monitoring has traditionally taken place in secondary care because of the need for a laboratory blood test, the international normalised ratio (INR). The development of reliable near patient testing (NPT) systems for INR estimation has facilitated devolution of testing to primary care. Patient self-management is a logical progression from the primary care model. ...
متن کاملPoint-of-care INR coagulometers for self-management of oral anticoagulation: primary care diagnostic technology update.
British Journal of General Practice, November 2012 e798 ADVANTAGES OVER EXISTING TECHNOLOGY Devices to measure international normalised ratio (INR) are intended for both professional use and patient self-testing to monitor oral anticoagulant therapy. INR is a standardised measurement of the prothrombin time, which is the time it takes blood to clot after addition of tissue factor. Point-of-care...
متن کاملIs home warfarin self-management effective? Results of the randomised Self-Management of Anticoagulation Research Trial.
AIMS The Warfarin Self-Management Anticoagulation Research Trial (Warfarin SMART) was designed to determine whether patients self-managing warfarin (PSM) using the CoaguChek device and a dosing algorithm developed for the trial could keep the INR (International Normalised Ratio) test in target range at least as often as patients managed by usual care by the family doctor or hospital clinic. M...
متن کاملValidation of the CoaguChek XS international normalised ratio point-of-care analyser in patients at Charlotte Maxeke Johannesburg Academic Hospital, South Africa.
BACKGROUND Measurement of the international normalised ratio (INR) is essential in the management of patients on long-term warfarin therapy. The CoaguChek XS portable coagulometer is a point-of-care test for INR measurement. It offers the advantage of improved patient accessibility, particularly in peripheral clinics. OBJECTIVES To evaluate the clinical utility of the CoaguChek XS for monitor...
متن کاملPoint of care monitoring of oral anticoagulant therapy in children: comparison of CoaguChek Plus and Thrombotest methods with venous international normalised ratio.
This paper reports the outcome of a research protocol aimed at optimising warfarin monitoring in a tertiary pediatric centre. The Thrombotest INR was the standard monitoring test employed to manage oral anticoagulant therapy in children at the Royal Children's Hospital (RCH), Melbourne. This study compares the results of this standard method to the novel CoaguChek INR monitor and the "gold stan...